Common IF Signature Problems

Region A of
Figure 8-4 on page 436
Suspect the first LC pole.
Region B of
Figure 8-4 on page 436
Suspect the third step-gain stage, the fine attenuator, or the fourth
LC-pole output amplifier.
Region C of Figure 8-4, "Detailed IF Adjust Signature (1)," has
no 15 dB step:
Suspect the reference 15 dB attenuator.
Region D of
Figure 8-4 on page 436
Suspect the second step-gain stage.
Entire signature noisy:
If the signature resembles
first step-gain stage or a break in the signal path in the input switch,
first crystal pole, or second crystal pole.
Correct shape but noisy:
If the signature resembles
crystal-pole output amplifier.
Amplitude of Region B of
than 12 dB:
Suspect the third step-gain stage output amplifier.
Region B of
Figure 8-12 on page 445
Suspect the fourth LC-pole output amplifier.
Chapter 8
is noisy:
is flat:
is flat:
Figure 8-9 on page
442, suspect a broken
Figure 8-10 on page
Figure 8-11 on page 444
is kinked:
IF Section
A5 IF Assembly
443, suspect the second
varies more
441