Safety inspection
3-
3
Safety inspection
3.1
Mechanical safety
SIM
Covers
•
Inspect the covers of the ARCOSKOP and the monitor trolley/MSS (monitor support
system) for mechanical damage and effective fastening.
SIM
Cable deflectors
•
Inspect the cable deflectors on the monitor trolley and replace them if necessary.
SIM
I.I. laser light localizer mechanism
•
Inspect the I.I. laser light localizer for mechanical damage.
•
Mount the I.I. laser light localizer on the I.I. and ensure proper locking and seating. Pay
special attention to the tension band and its closure.
SIM
I.I. laser light localizer function
•
Perform maintenance on the I.I. laser light localizer according to Installation and Set-
ting Instructions RXR2-120.033.03...
SIM
Laser light localizer mechanism
•
Inspect the I.I. attachment of the laser light localizer for mechanical damage.
•
Install the I.I. attachment of the laser light localizer on the I.I. and ensure proper locking
and seating.
•
Check the I.I. ring for damage.
SIM
Brakes
•
Check the gastro foot switch and brake function.
•
Use the spring balance to check the braking forces for the orbital movement, angula-
tion movement, and rotary joints.
SIM
C-arm
•
Perform all C-arm movements, checking for slackness of the bearings and bearing
noises.
•
Orbital movement: When the orbital brake is released, the C-arm may independently
move a maximum of 5°. Check the servo assistance of the orbital movement for proper
operation. Check whether the orbital movement is running smoothly.
SIM
Stand
•
Check the corrugated hoses on the ARCOSKOP stand for damage.
•
Check the stand joint covers, closing caps of the stand members, and the covers of the
ceiling installation for proper fastening.
SIM
Remote control panel
•
Check the corrugated hose for mechanical damage.
•
Ensure proper functioning of the locking screw for the remote control panel.
•
Check the remote control panel for mechanical damage.
SIM
E-box
Siemens AG
Medical Solutions
SPR2-250.831.01.03.02
02.05
Page 15 of 26
CS PS 24
15