Attaching the swivel-arm covers

Attaching the swivel-arm covers

Arranging the swivel-arm system

1
Install object table, no compression force (swivel-arm system still upside down).
2
Rotate the swivel-arm system to 0 degrees by pressing one of the switches for clock-
wise rotation on the supporting arm.
CAUTION!
Do not perform any vertical adjustment of the swivel-arm system, before the protective
strips have been removed from the metal curtain
3
System and mains "OFF".
4
Remove the swivel-arm system transport safeguard, see section 3.

Connecting the cables to control-button board and patient handles

Figure 1 Connecting the cables
Before installing side covers, cables are to be connected.
1
Connect cables X807 (1) to control-button circuit boards D807 (2).
2
Connect ground wires to patient handles (3).
Siemens-Elema AB
Solna, Sweden
2
2
SPS-UD
RX B7-120.033.01.10.02
4
3
1
1
3
4
05.98
8-1
MAMMOMAT 300
65 02 186