Maintenance
Exchanging the Rotor Seal
5
Install the new rotor seal and isolation seal. Ensure the
metal spring inside the isolation seal faces towards the
valve body.
7
Install stator head. Tighten the bolts alternately two turns
at a time until the stator head is secure.
118
6
Install the stator ring with the short of the two pins facing
towards you at the 12 O'Clock position. Ensure the ring
sits flat on the valve body.
8
Reconnect the pump capillaries to the valve ports (see
Hydraulic Connections).
Agilent 1260 Infinity Standard Autosampler User Manual