8
Maintenance

Exchanging the Rotor Seal

Exchanging the Rotor Seal

When
Poor injection-volume reproducibility
Leaking injection valve
Tools required
Description
Wrench, 1/4 inch (supplied in HPLC Tool-Kit)
Hexagonal key, 9/64 inch (supplied in HPLC Tool-Kit)
Parts required
#
1
1
1
Preparations
• Remove front cover.
• Remove the leak tubing (if necessary).
Removing the stator head
C A U T I O N
The stator face is held in place by the stator head. When you remove the stator head,
the stator face can fall out of the valve.
116
p/n
0100-1853
0100-1849
0101-1416
Carefully handle the valve to prevent damage to the stator face
Description
Rotor seal (Vespel)
Rotor seal (Tefzel)
Rotor seal (PEEK)
Agilent 1260 Infinity Standard Autosampler User Manual