66 | DX-D 600 | Getting Started with DX-D 600
Step 5: Execute the exposure
In the operator room:
Press the exposure button to execute the exposure.
The exposure is sent to the NX workstation.
In the operator room at the NX workstation:
• The image is acquired from the DR Detector and displayed in the
thumbnail.
• If collimation is applied, the image is automatically cropped at the
collimation borders.
• The actual X-Ray exposure parameters are sent back from the console to
the NX workstation and are shown in the Image Detail pane.
• The actual X-Ray exposure parameters and the Exposure Index (EI) value
on the NX workstation can be used to monitor the performance of the
Automatic Exposure Control of the X-Ray system.
0240H EN 20181119 0940
WARNING:
The radiation indicator on the control console lights up during
exposure release.
WARNING:
Do not select another thumbnail until the preview image is
visible in the active thumbnail.