Starting DX-D 600

To start DX-D 600:
1. Switch on the DR Generator Sync Box.
2. Switch on the X-Ray generator control in the operator room.
The X-Ray generator and the wired DR Detectors are powered on.
3. Switch on the X-Ray system using the button on the X-Ray system
electrical cabinet in the examination room.
The X-Ray system and the control panel touch screen console are powered
on.
4. Start NX.
The NX application and the software console are available on the NX
workstation.
For detailed information about starting up NX, refer to the NX User
Manual, document 4420.
5. In a configuration with a wireless DR Detector, power on the DR Detector:
• attach a fully charged battery pack to the DR Detector.
• turn on the DR Detector.
• if needed, register the DR Detector to the NX workstation.
For detailed information about starting up the DR Detector, refer to the DR
Detector User Manual.
Note: Allow the DR Detector to warm up before the DX-D 600 is
used for clinical purposes. The warming-up time starts as soon as
the DR Detector has been powered on and the NX workstation is
running. To check if a warming-up time is required, refer to the DR
Detector User Manual.
DX-D 600 | Getting Started with DX-D 600 | 59
0240H EN 20181119 0940