Dx-D 600 Technical Data - AGFA DX-D 600 User Manual

Hide thumbs Also See for DX-D 600:
Table of Contents

Advertisement

86 | DX-D 600 | Technical Data

DX-D 600 Technical Data

The technical data are provided in this chapter or in the User Manual of the
component.
Table 1: Environmental conditions for the X-ray system
Environmental Conditions (during storage and transport)
Temperature (ambient)
Humidity (non condensing)
Atmospheric pressure
Environmental Conditions (during normal operation)
Temperature (ambient)
Humidity (non condensing)
Atmospheric pressure
For overall system environmental conditions, the environmental conditions of
the DR Detector or image plate should be taken into account. Refer to the
related User Manual for environmental conditions for the DR Detector or
image plate. When using the DR Detector or image plate inside the bucky, take
into account that the temperature inside the bucky can be up to 5°C higher
than the temperature in the X-ray room.
0240H EN 20181119 0940
between -20 and 70 degrees Celsius
between 10 and 70 % relative humidity
between 50 and 106 kPa
between 10 and 35 degrees Celsius
between 30 and 75 % relative humidity
between 70 and 106 kPa

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dx-d 600 generic5430/1105430/120

Table of Contents