Problem With Automatic Movements Of The X-Ray System - AGFA DX-D 600 User Manual

Hide thumbs Also See for DX-D 600:
Table of Contents

Advertisement

Problem with automatic movements of the X-ray
system
Details
Cause
Brief Solution
Related Links
Stopping DX-D 600
Starting DX-D 600
The automatic centering or auto-tracking is not working
as expected.
-
Use the emergency stop button to reset the system:
1. Press the emergency stop button.
2. Release the emergency stop button.
A beep indicates that the system is reset.
3. A message may be displayed on the software console
to confirm position synchronization.
4. On the NX workstation, select a new thumbnail.
5. Press and hold the AUTOCENTER button on the au-
tomatic positioning control box.
If the automatic positioning is not restored, perform a
shutdown and startup of the whole system.
on page 79
on page 59
DX-D 600 | Problem solving | 81
0240H EN 20181119 0940