Electromagnetic Immunity - AGFA DR 10e User Manual

Table of Contents

Advertisement

Electromagnetic immunity

This DR Detector is intended for operation in the electromagnetic
environment given below. The user of the DR Detector should ensure that it is
used in such an environment.
Immunity Test
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short interrup-
tions and voltage
variations on
power supply in-
put lines.
IEC 61000-4-11
DR 10e, DR 14e, DR 17e | Remarks for HF-emission and immunity | 127
IEC 60601-1-2
Test Level
± 8 kV contact
± 15 kV air
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
0% U
for 0.5 cy-
T
cle
0% U
for 1 cycle
T
70% U
(30% dip
T
in U
) for 25 cy-
T
cles
0% U
for 5 s
T
Compliance
Electromagnetic en-
Level
vironment - guid-
± 8 kV con-
Floors should be
tact
wood, concrete or ce-
ramic tile. If floors
± 15 kV air
are covered with syn-
thetic material, the
relative humidity
should be at least
30%.
± 2 kV for
Mains power quality
power supply
should be that of a
lines
typical commercial or
hospital environ-
± 1 kV for in-
ment.
put/output
lines
± 1 kV
Mains power quality
line(s) to
should be that of a
line(s)
typical commercial or
hospital environ-
± 2 kV
ment.
line(s) to
earth
0% U
for 0.5
Mains power quality
T
cycle
should be that of a
typical commercial or
0% U
for 1
T
hospital environ-
cycle
ment. If the user of
70% U
the device requires
T
(30% dip in
continued operation
U
) for 25 cy-
during power mains
T
cles
interruptions, it is
recommended that
0% U
for 5 s
T
the device be pow-
ered from an uninter-
ruptible power supply
or a battery.
0370C EN 20200228 1449