Starting the DR Detector (wired configuration)

To start the DR Detector:
1. Atach the cover plate of the battery bay if no battery is attached.
2
Align the cover plate according to the guide marks. Insert the cover
1.
plate fully.
Push down the cover plate.
2.
Figure 21: Attach the cover plate of the battery bay
The cover plate is locked in position automatically.
2. Plug the power cable of the DR Detector power box into a mains power
socket.
The power status light on the front panel of the DR Detector power box
lights up.
3. Connect the DR Detector cable.
Push the connector of the cable straight into the connector slot of the DR
Detector.
The detector starts up. The power indicator lights up green.
4. Check the DR Detector status icon on the DR Detector Switch. If the
displayed status is error and the detector is shared between NX
workstations, it may still be connected to another NX workstation. In that
case, register the DR Detector to the NX workstation.
The connection indicator is lit. The DR Detector is ready.
Before exposure make sure to check the equipment daily and confirm that it
works properly.
Note: Before operating the detector, start up the NX workstation.