Artifact In Dr Detector Images; Dr Detector Status Is Not Changing To Ready For Exposure - AGFA DR 10e User Manual

Table of Contents

Advertisement

Artifact in DR Detector images

Details
Cause
Brief Solution
DR Detector status is not changing to ready for
exposure
Details
Cause
Brief Solution
DR 10e, DR 14e, DR 17e | Problem solving | 111
An artifact is visible in the images produced by a DR
Detector.
Exposure conditions have changed significantly since
latest calibration.
Perform calibration of the DR Detector.
For details, refer to the DX-D DR Detector Calibration
Key User Manual (0134).
Select the thumbnail for an exposure in the Image
Overview pane of the Examination window.
The DR Detector Switch shows the active DR Detector
and shows its status.
The DR Detector status is not changing to ready for ex-
posure.
The DR Detector is performing an internal process.
Wait at least 2 minutes and select the thumbnail again.
0370C EN 20200228 1449