126 | DR 10e, DR 14e, DR 17e | Remarks for HF-emission and immunity

Electromagnetic emissions

This DR Detector has been tested for the electromagnetic environment as
described below.
The user of the DR Detector should ensure that it is used in such an
environment.
Nevertheless the HF-emission and immunity can be influenced by connected
data cables depending on length and the manner of installation.
Emissions test
RF emissions in ac-
cordance with CISPR
11
RF emissions in ac-
cordance with CISPR
11
Harmonic emissions
in accordance with
IEC 61000-3-2
Voltage fluctuations /
flicker emissions in
accordance with
IEC 61000-3-3
0370C EN 20200228 1449
Compli-
ance
Group 1 This DR Detector uses RF energy only for
Class B
Com-
plies
Com-
plies
Note: IEC 61000-3-2 and IEC 61000-3-3 are applicable only to
the devices with a rated voltage of 220 V AC or higher.
Electromagnetic Environment Guidelines
its internal function.
Therefore, its RF emissions are very low
and are not likely to cause any interfer-
ence in nearby electronic equipment.
The DR Detector is suitable for use in all
establishments, including domestic estab-
lishments and those directly connected to
the public low-voltage power supply net-
work that supplies buildings used for do-
mestic purposes.