Recommended Separation Distance - AGFA DR 10e User Manual

Table of Contents

Advertisement

Recommended separation distance

This device is intended for operation in an electromagnetic environment in
which the radiated high frequency disturbance variables are monitored. The
user of the device can help to prevent electromagnetic disruptions by
maintaining the minimum distances between portable and mobile high
frequency communication equipment (transmitters) and the device as
recommended below, in accordance with the maximum output power of the
communications equipment.
Recommended Protective Distances between Portable and Mobile High Fre-
quency Communication Equipment and the Device
Rated Power of
the Transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the rec-
ommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maxi-
mum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
REMARK 1: At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
REMARK 2: These Guidelines may not be relevant in all situations. The dis-
persion of electromagnetic waves is influenced by absorption and reflections
from buildings, objects and people.
DR 10e, DR 14e, DR 17e | Remarks for HF-emission and immunity | 131
Protective Distance in accordance with Transmission
Frequency
m
150 kHz to 80
MHz
d = 1.2
0.12
0.38
1.2
3.8
12
80 MHz to 800
MHz
d = 1.2
0.12
0.38
1.2
3.8
12
0370C EN 20200228 1449
800 MHz to 2.5
GHz
d = 2.3
0.23
0.73
2.3
7.3
23

Hide quick links: