86 | DX-D 60C, DX-D 60G | Remarks for HF-emission and immunity

Electromagnetic immunity

This DR Detector is intended for operation in the electromagnetic
environment given below. The user of the DR Detector should ensure that it is
used in such an environment.
Resistance to
Jamming Test
Discharge of
static electrici-
ty in accord-
ance with
IEC 61000-4-2
Fast transient
electrical dis-
turbance vari-
ables / bursts
in accordance
with IEC
61000-4-4
Impulse vol-
tages (surges)
in accordance
with IEC
61000-4-5
Voltage break-
throughs,
short term in-
terruptions
and variations
in the voltage
supplied in ac-
cordance with
IEC
61000-4-11
0294B EN 20190221 1131
IEC 60601
Level of
Test Level
Agreement
± 8 kV contact
± 8 contact
discharge
discharge
± 15 kV air
± 15 kV air
discharge
discharge
± 2 kV for net-
± 2 kV for net-
work leads
work leads
± 1 kV for en-
± 1 kV for en-
try and outlet
try and outlet
leads
leads
± 1 kV push-
± 1 kV push-
pull voltage
pull voltage
± 2 kV com-
± 2 kV com-
mon mode
mon mode
voltage
voltage
100% reduc-
100% reduc-
tion for 0.5 cy-
tion for 0.5 cy-
cle at 0, 45, 90,
cle at 0, 45, 90,
135, 180, 225,
135, 180, 225,
270 and 315
270 and 315
degrees
degrees
100% reduc-
100% reduc-
tion for 1 cycle
tion for 1 cycle
30% reduction
30% reduction
for 25/30 cy-
for 25/30 cy-
cles at 0 degree
cles at 0 degree
100% reduc-
100% reduc-
tion for
tion for
250/300 cycles
250/300 cycles
(5 sec.)
(5 sec.)
Electromagnetic Envi-
ronment Guidelines
Floors should consist of
wood, concrete or ce-
ramic tiles. The relative
humidity must be at least
30%, if the floor is made
of synthetic material.
The quality of the volt-
age supplied should cor-
respond to a typical com-
mercial or clinical envi-
ronment.
The quality of the volt-
age supplied should cor-
respond to that of a typi-
cal commercial or clini-
cal environment.
The quality of the volt-
age supply should corre-
spond to that of a typical
commercial or clinical
environment. If the user
wants the DR Detector to
work continuously, even
when the energy supply
is interrupted, it is rec-
ommended to use an en-
ergy supply free of inter-
ruptions or a battery.