Precautions on EMC

Medical electrical equipment needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC information
provided in the manual.
Portable and mobile RF communications equipment can affect medical
electrical equipment.
DR 10e, DR 14e, DR 17e | Remarks for HF-emission and immunity | 123
WARNING:
The DR Detector should not be used adjacent to or stacked with
other equipment and if adjacent or stacked use is necessary, the
DR Detector should be observed to verify normal operation in
the configuration in which it will be used.
WARNING:
Avoid to place the DR Detector too close to life supporting
devices. Keep a minimal distance of 26 cm between the DR
Detector and life supporting devices.
WARNING:
Do not place devices generating electromagnetic wave near this
equipment.
WARNING:
If devices other than those specified are connected,
predetermined EMC performance cannot be guaranteed.
WARNING:
Do not use mobile RF communications equipment within 30 cm
(11.8 inch) of this equipment.
WARNING:
The DR detectors might be interfered with by other equipment,
even if that other equipment complies with CISPR emission
requirements.
WARNING:
Do not use this equipment near devices that generate strong
electromagnetic waves, such as thermotherapy devices or HF
surgical equipment.
WARNING:
If this equipment is used near commercially available electronic
devices, such as mobile phones, laptop computers or home
appliances, which generate electromagnetic waves, this
0370C EN 20200228 1449