74 | DR 10e, DR 14e, DR 17e | Getting started

Starting the DR Detector (wireless configuration)

To start the DR Detector:
1. Fully charge the battery.
Charge the battery on the day of examination or on the previous day.
2. Remove the cover plate of the battery bay.
a) Push the lock lever at the side with the dot.
b) Slide the lock lever in the direction of the arrow.
0370C EN 20200228 1449
CAUTION:
Do not use the battery pack as a power source for equipment
other than DR 10e, DR 14e or DR 17e detectors. Be sure to use
only the dedicated battery pack for the DR 10e, DR 14e or DR
17e detector.
CAUTION:
Monitor the battery status. If the charging level of the battery is
low, the battery has to be charged or replaced by another
battery.
Note: Before operating the detector, start up the NX workstation.
Note: The battery slowly discharges even if not in use. The
battery pack may have expired if it discharges immediately
after being fully charged. You can purchase an optional battery
pack to replace an exhausted one.