WARNING
Discontinue using or replace this mask if the patient has ANY adverse reaction to the
use of the mask. Consult your physician or sleep therapist.
Magnets are used in the lower headgear straps and the frame of the mask. Ensure the
headgear and frame is kept at least 50 mm away from any active medical implant (eg,
pacemaker or defibrillator) to avoid possible effects from localized magnetic fields.
The magnetic field strength is less than 400 mT.
The mask is not intended to be used simultaneously with nebuliser medications that
are in the air path of the mask/tube.
If there is any visible deterioration of a mask component (cracking, crazing, tears etc.)
the component should be discarded and replaced.
CAUTION
Always follow cleaning instructions and only use a mild liquid detergent. Some
cleaning products may damage the mask, its parts and their function, or leave harmful
residual vapours. Do not use a dish washer or washing machine to clean the mask.
When fitting the mask, do not overtighten the headgear as this may lead to skin
redness or sores around the mask cushion.
As with all masks, some rebreathing may occur at low pressures.
Using a mask may cause tooth, gum or jaw soreness or aggravate an existing dental
condition. If symptoms occur, consult your physician or dentist.
Do not iron the headgear as the material is heat sensitive and will be damaged.

Technical specifications

Mask setting options: For AirSense, AirCurve or S9: Select 'Full Face'.
Compatible devices: For a full list of compatible devices for this mask, see the
Mask/Device Compatibility List at ResMed.com/downloads/masks.
Pressure-flow curve
Therapy pressure: 4 to 30 cm H
6
0
2
Pressure
Flow
(cm H
O)
(L/min)
2
4
21
11
36
17
47
24
57
30
65