Disinfection and sterilization guide
This guide is intended for multipatient use of the AirFit
Guide for cleaning instructions. This guide describes ResMed's recommended and validated procedures for cleaning, disinfection and sterilization of the mask in accordance with ISO17664.
167°F (75°C)-30 min;
AirFit F20 / AirTouch F20
176°F (80°C)-10 min;
194°F (90°C)-1 min
mask component
1
• AirFit F20 cushion
• AirTouch F20 cushion
3
• Elbow
• Frame
• Headgear
1
This mask may not be available in all regions. For full details regarding the correct use of this mask, please refer to the specific User Guide. For a list of available replacement parts for each mask system, check the Components Card on www.resmed.com.
2
If a healthcare facility requires an additional disinfection or sterilization cycle after reassembly, the number of validated cycles must be halved.
3
The AirTouch F20 cushion is intended for single-patient reuse in both home and hospital/institutional environments and recommended to be replaced monthly.
4
Not validated.
A
F20 / AirTouch
F20 ResMed full face mask in a sleep lab, clinic or hospital. If you use the mask as a single user in the home, refer to the User
High level thermal disinfection
200°F (93°C)-10 min
B
C
1
2
3
4
High level chemical disinfection
CIDEX
OPA
STERRAD
Ortho-phthalaldehyde 0.55%
A
D
1
2
3
4
B
C
D
5
6
7
A+B+C
5 6
7
A+B+C+D
Sterilization
Validated number of cycles
100S
STERRAD
NX
Standard or
Advanced cycle
Elbow
Valve (AAV)
Vent
Side buttons
Swivel
Frame
Cushion
Headgear
Magnetic clips
Lower headgear straps
Upper headgear straps
Frame system
Complete system
Clinical use only
2
30
N/V
4
30
30
30
English
1

Advertisement

loading

Summary of Contents for ResMed AirFit F20

  • Page 1 This mask may not be available in all regions. For full details regarding the correct use of this mask, please refer to the specific User Guide. For a list of available replacement parts for each mask system, check the Components Card on www.resmed.com.
  • Page 2 Disinfection 1. Using a hot water disinfection system, soak the disinfectable mask components using a temperature-time combination: • AirFit F20 cushion/elbow/frame: 167°F (75°C) for 30 minutes; 176°F (80°C) for 10 minutes; 194°F (90°C) for 1 minute; 200°F (93°C) for 10 minutes 200°F (93°C)
  • Page 3 Disassemble the mask according to the instructions in the User Guide. Cleaning Alconox AirFit F20 cushion 1. Make a solution of Alconox by diluting with drinking quality water at 1%, ie, 10 g per liter, at 68-77°F (20-25°C), according to manufacturer’s instructions. 2. Soak the component in the solution.
  • Page 4 Cleaning Alconox AirFit F20 cushion 1. Make a solution of Alconox by diluting with drinking quality water at 1%, ie, 10 g per liter, at 68-77°F (20-25°C), according to manufacturer’s instructions. 2. Soak the component in the solution for 1 minute.
  • Page 5: General Warnings And Cautions

    Do not iron the headgear as the material is heat sensitive and will be damaged. ResMed Ltd See ResMed.com for other ResMed locations worldwide. AirFit and AirTouch are trademarks and/or registered trademarks of the ResMed family of companies. CIDEX and STERRAD are trademarks of Johnson & Johnson. 1 Elizabeth Macarthur Drive Alconox is a trademark of Alconox Inc.