Technical Specifications - ResMed Mirage Quattro User Manual

Full face mask
Hide thumbs Also See for Mirage Quattro:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 4

Technical Specifications

Note: The mask system does not contain latex, PVC or DEHP materials.
Pressure–Flow Curve
70.0
60.0
50.0
0.0
30.0
20.0
10.0
0.0
2
6
8
10
Mask Pressure (cm H
Note: As a result of manufacturing variations, the vent flow rate may vary.
Dead Space Information
Dead space: the empty volume of the mask up to the swivel.
The dead space of the mask varies according to cushion size. It is 242 mL for the large
size.
Standards
CE designation in accordance with EC directive 93/42/EEC, class IIa.
ISO 14971, ISO 13485, ISO 10993-1 and ISO 17510-2.
Therapy Pressure
4 to 20 cm H
O
2
Resistance
Drop in pressure measured (nominal)
at 50 L/min: 0.5 cm H
at 100 L/min: 1.6 cm H
Sound
The A-weighted sound power of the mask is 28 dBA.
Environmental Conditions
Operating temperature: +5°C to +40°C
Operating humidity: 10%–95% relative humidity non-condensing
Storage and transport: -20°C to +60°C
Storage and transport humidity: 10%–95% relative humidity non-condensing
12
1
16
18
20
22
2
26
O)
2
O
2
O
2
Pressure
(cm H
O)
2
4
8
12
16
20
28
30
7
Flow
(± 6.0 L/min)
22
32
41
48
54

Advertisement

Table of Contents
loading

Table of Contents