Technical Specifications - ResMed AirFit P10 User Manual

Hide thumbs Also See for AirFit P10:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 5
Problem/possible cause
Mask too noisy
System is assembled incorrectly. Disassemble the mask, then reassemble according to
System is leaking air.

Technical specifications

Pressure-flow curve
Dead space information
Therapy pressure
Resistance
Sound
Environmental
conditions
Solution
the instructions.
Refit or reposition your mask. Adjust your headgear.
Check your mask is assembled correctly.
The mask contains passive venting to protect against
rebreathing. As a result of manufacturing variations, the vent
flow rate may vary.
Physical dead space is the empty volume of the mask to the
end of the swivel. Using the large cushions it is 123 mL.
4 to 20 cm H
O
2
Drop in pressure measured (nominal)
at 50 L/min: 0.4 cm H
at 100 L/min: 1.4 cm H
DECLARED DUAL-NUMBER NOISE EMISSION VALUES in
accordance with ISO 4871. The A-weighted sound power
level of the mask is 21 dBA, with uncertainty of 3 dBA. The
A-weighted sound pressure level of the mask at a distance
of 1 m is 13 dBA, with uncertainty of 3 dBA.
Operating temperature: +5°C to +40°C
Operating humidity: 15% to 95% non-condensing
Storage and transport temperature: -20°C to +60°C
Storage and transport humidity: up to 95% relative humidity
non-condensing
O
2
O
2
Pressure
Flow
(cm H2O)
(L/min)
4
20
8
29
12
37
16
43
20
49
English
7