Technical Specifications - ResMed S9 AutoSet Information Manual

Positive airway pressure devices
Hide thumbs Also See for S9 AutoSet:

Advertisement

Problem/Possible cause
Displays error message: High leak, please check system setup and all connections
There is excessive leak.
The following message is displayed on the LCD after you try to update settings or copy data to
the SD card: Card error, please remove SD card and contact service provider
SD card is not inserted correctly.
You may have removed the SD card
before settings were copied to the CPAP
device.
The following message is NOT displayed on the LCD after you try to update the settings using
the SD card: Settings updated successfully, press any key
The settings were not updated.

Technical specifications

Operating pressure range
Maximum single fault steady state
pressure
Pressure measurement tolerance
Flow measurement tolerance
DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance with ISO 4871:1996
Sound pressure level
Sound power level
Nominal dimensions (L x W x H)
Weight
4
Solution
Adjust position of mask and headgear.
Air tubing is not connected properly. Connect firmly at
both ends.
Ensure that the SD card is inserted correctly.
Reinsert the SD card and wait for the home screen or the
Settings updated successfully, press any key
message to appear on the LCD.
Note: This message only appears once. If you re-insert
the SD card after you have updated your settings, the
message will not be re-displayed.
Contact your clinician/service provider immediately.
4 to 20 cm H
O
2
30 cm H
O
2
±0.5 cm H
O ± 4% of the measured reading
2
±6 L/min or 10% of reading, whichever is greater, at 0 to
150 L/min positive flow
24 dBA as measured according to ISO 17510-1:2002
26 dBA with uncertainty of 2 dBA as measured according to
EN ISO 17510-1:2009
34 dBA with uncertainty of 2 dBA as measured according to
EN ISO 17510-1:2009
153 mm x 140 mm x 86 mm
835 g