Technical specifications

90W power supply unit
AC input range:
DC output:
Typical power consumption:
Peak power consumption:
Environmental conditions
Operating temperature:
Operating humidity:
Operating altitude:
Storage and transport temperature:
Storage and transport humidity:
Electromagnetic compatibility
The AirCurve 10 complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-
2:2007, for residential, commercial and light industry environments. It is recommended that mobile communication devices
are kept at least 1 m away from the device.
Information regarding the electromagnetic emissions and immunity of this ResMed device can be found on
www.resmed.com, on the Products page under Service and Support.
IEC 60601-1:2005 classification
Class II (double insulation), Type BF, Ingress protection IP22.
Sensors
Pressure sensor:
Flow sensor:
Maximum single fault steady pressure
Device will shut down in the presence of a single fault if the steady state pressure exceeds:
30 cm H
O for more than 6 sec or 40 cm H
2
Sound
Pressure level measured according to ISO 17510-1:2007 (CPAP mode):
SlimLine:
Standard:
Power level measured according to ISO 17510-1:2007 (CPAP mode):
SlimLine:
Standard:
Declared dual-number noise emission values in accordance with ISO 4871:1996.
Physical - device and water tub
Dimensions (H x W x D):
Air outlet (complies with ISO 5356-1:2004):
Weight (device and standard water tub):
Weight (device and cleanable water tub):
Housing construction:
16
100–240V, 50–60Hz 1.0–1.5A, Class II
115V, 400Hz 1.5A, Class II (nominal for aircraft use)
24V
53W (57VA)
104W (108VA)
+41°F to +95°F (+5°C to +35°C)
Note: The air flow for breathing produced by this therapy
device can be higher than the temperature of the room.
Under extreme ambient temperature conditions
(104°F/40°C) the device remains safe.
10 to 95% relative humidity, non-condensing
Sea level to 8,500' (2,591 m); air pressure range 1013 hPa to
738 hPa
-4°F to +140°F (-20°C to +60°C)
5 to 95% relative humidity, non-condensing
Internally located at device outlet, analog gauge pressure
type, -5 to +45 cm H
Internally located at device inlet, digital mass flow type, -70
to +180 L/min
O for more than 1 sec.
2
26.6 dBA with uncertainty of 2 dBA
26.6 dBA with uncertainty of 2 dBA
34.6 dBA with uncertainty of 2 dBA
34.6 dBA with uncertainty of 2 dBA
4.57" x 10.04" x 5.91"
(116 mm x 255 mm x 150 mm)
22 mm
44 oz (1248 g)
44 oz (1248 g)
Flame retardant engineering thermoplastic
3.75A
O
2