Colorimetric analyzer for total nitrogen (32 pages)
Summary of Contents for Endress+Hauser Liquiline System CA80SI
Page 1
These instructions are Brief Operating Instructions; they are not a substitute for the Operating Instructions pertaining to the device. Detailed information on the device can be found in the Operating Instructions and in the other documentation available at: • www.endress.com/device-viewer • Smart phone/tablet: Endress+Hauser Operations App...
Page 2
Liquiline System CA80SI Order code: XXXXX-XXXXXX Ser. no.: XXXXXXXXXXXX Ext. ord. cd.: XXX.XXXX.XX Serial number www.endress.com/deviceviewer Endress+Hauser Operations App A0023555 Endress+Hauser...
About this document Liquiline System CA80SI About this document Warnings Structure of information Meaning This symbol alerts you to a dangerous situation. DANGER Failure to avoid the dangerous situation will result in a fatal or serious injury. Causes (/consequences) If necessary, Consequences of non- compliance (if applicable) ‣...
Liquiline System CA80SI About this document Symbols at the device Symbol Meaning Reference to device documentation Caution: Hazardous voltage Documentation The following instructions complement these Brief Operating Instructions and are available on the product pages on the Internet: • Operating Instructions Liquiline System CA80 –...
Designated use The Liquiline System CA80SI is a wet-chemical analyzer for almost continuous determination of the silica concentration in ultrapure water and boiler feedwater. The analyzer is designed for use in the following applications: •...
Liquiline System CA80SI Basic safety instructions Ensure that electrical cables and hose connections are undamaged. Do not operate damaged products, and protect them against unintentional operation. Label damaged products as defective. During operation: ‣ If faults cannot be rectified: products must be taken out of service and protected against unintentional operation.
Incoming acceptance and product identification Liquiline System CA80SI Incoming acceptance and product identification Incoming acceptance Verify that the packaging is undamaged. Notify the supplier of any damage to the packaging. Keep the damaged packaging until the issue has been resolved.
Liquiline System CA80SI Incoming acceptance and product identification 3.2.2 Product identification Product page www.endress.com/ca80si Interpreting the order code The order code and serial number of your product can be found in the following locations: • On the nameplate • In the delivery papers Obtaining information on the product Open the product website.
Installation Liquiline System CA80SI Accessories (dependent on number of channels): 1-channel 2-channel 4-channel 6-channel Filters and 1 filter, 1 pressure 2 filters, 2 pressure Panel with 4 pre-installed Panel with 6 pre-installed pressure relief relief valve with relief valves with...
Page 11
Liquiline System CA80SI Installation 4.1.1 Dimensions 463 (18.23) 530 (20.87) A0028820 1 Liquiline System CA80 closed version, dimensions in mm (in) 417 (16.42) 530 (20.87) A0030419 2 Liquiline System CA80 open version, dimensions in mm (in) Endress+Hauser...
Page 12
Installation Liquiline System CA80SI 654 (25.74) A0028821 3 Liquiline System CA80 with base, dimensions in mm (in) Endress+Hauser...
Page 13
Liquiline System CA80SI Installation (1.61) (2.20) A0036334 4 CA80SI 1-/2-channel version: Filter (left), pressure-reducing valve (right), dimensions in mm (in) 40 (1.57) 50 (1.97) A0036665 5 Dimensions of angle bracket for filter Fasteners (2 x M5) 54 (2.13) 63,5 (2.5)
Page 14
Installation Liquiline System CA80SI 8 (0.31) 355 (13.98) 335 (13.19) (3.74) A0036389 7 CA80SI 4-/6-channel version: Panel with pressure-reducing valves and filters, dimensions in mm (in) 8 (0.31) 255 (10.04) (2.91) (9.25) A0036390 8 CA80SI 4-/6-channel version: Panel with sample channel switching, dimensions in mm (in) 4.1.2...
Page 15
Liquiline System CA80SI Installation 4.1.3 Spacing requirements when mounting Spacing required for installing analyzer ≥330(12.99) ≤116 ° A0036774 A0036775 9 Minimum spacing required for mounting. 10 Maximum opening angle Engineering unit mm (in). Spacing required for installing wall-mount version Ø...
Installation Liquiline System CA80SI Mounting the analyzer on a wall CAUTION Incorrect installation can cause injury and damage the device ‣ If mounting on a wall, check that the analyzer is fully hooked into the wall holder unit at the top and bottom, and secure the analyzer to the upper wall holder unit using the securing screw.
Liquiline System CA80SI Installation Hook the analyzer into the wall holder unit. Secure the two top parts of the wall holder unit with the screw supplied. Installing version with analyzer stand CAUTION Incorrect installation can cause injury and damage the device ‣...
Installation Liquiline System CA80SI A0036785 15 Securing the base Screw the base to the ground. With 2 people, lift the analyzer and fit it on the base. Use the recessed grips. Secure the base to the analyzer using the 6 screws supplied.
Page 19
Liquiline System CA80SI Installation 1-/2-channel device: Installation area for pressure relief valve and filter L < 3 m L < 0,5 m (9.84 ) (1.64 ) A0036573 16 Permitted installation area, engineering unit m (ft) A0036671 17 Installing angle brackets for pressure relief valve and filter...
Installation Liquiline System CA80SI A0036743 Correct flow direction of pressure relief valve (indicated by triangle on pressure relief valve) Correct flow direction of filter (indicated by arrow on filter) Cut hose piece (polyurethane hose) to the required length (< 0.5 m (1.64 ft)).
Page 21
Liquiline System CA80SI Installation 4-/6-channel device: Installation area for sample channel switching and panel with pressure relief valves and filters L < 1m D + H < 3m (3.28 ) (9.84 ) A0036574 19 Permitted installation area, can be installed to the left or right of the analyzer, engineering unit m...
Installation Liquiline System CA80SI D + H < 3m (9.84 ) A0036667 20 Permitted installation area, engineering unit m (ft) 4-/6-channel version: Installing panel with sample channel switching Mounting materials are not supplied. ‣ The mounting materials must be provided by the customer onsite.
Liquiline System CA80SI Electrical connection A0036341 21 Panel with sample channel switching Post-installation check After mounting, check all the connections to ensure they are secure. Electrical connection WARNING Device is live! Incorrect connection may result in injury or death! ‣...
Electrical connection Liquiline System CA80SI Connecting the analyzer NOTICE The device does not have a power switch ‣ You must install the device near (distance < 3 m (10 ft)) an easily accessible and fused plug socket so that it can be disconnected from the power supply.
Liquiline System CA80SI Electrical connection Replace the cable glands in order versions G' and NPT: Replace the already installed M-thread cable glands with the enclosed G’- or NPT cable glands. This does not affect the M32 hose glands. For cabinet versions, the cable length is approx. 4.3 m (14.1 ft) from the base of the housing.
Electrical connection Liquiline System CA80SI Individual types of protection permitted for this product (integrity (IP), electrical safety, EMC interference immunity, Ex protection) can no longer be guaranteed if, for example: • Covers are left off • Different power units to the ones supplied are used •...
Liquiline System CA80SI Operation options Operation options Structure and function of the operating menu Analyzer_CA80 09:11:05 01.03.2018 SP1: Silica g/l SiO2 MENU MODE DIAG HOLD MODE DIAG A0036371 A0036786-EN 22 Overview of operation 23 Display (example) Menu path and/or device designation...
Commissioning Liquiline System CA80SI Commissioning Preparatory steps Before connecting the device to the power supply, connect the liquid-bearing lines and visually inspect the hoses. The reagents used can pose a health hazard. Pay attention to the information in the reagent safety data sheets.
Page 29
Liquiline System CA80SI Commissioning 7.1.1 Hose connection diagram The hose connection diagram is also located • on the inside of the analyzer door • in the Operating Instructions for the analyzer (→ 5) RN RK RB SP1 SP2 D Sample drain / Probenablauf / Reagent RB / Reagenz RB / Réactif RB...
Commissioning Liquiline System CA80SI 4-/6-channel version A constant and sufficient supply of sample must be ensured at the installation location. If a sample channel is not in use, leave the red drain plug in the valve. Remove the drain plug from the sample channels.
Liquiline System CA80SI Commissioning Configuring the measuring device 7.4.1 Basic setup analyzer Making basic settings Switch to the Setup/Basic setup analyzer menu. Make the following settings. Device tag: Give your device any name of your choice (max. 32 characters).
Commissioning Liquiline System CA80SI The device is now ready to measure. At the start of the initial measurement, the reagent dispensers are inflated completely and emptied. This is to guarantee the measuring performance from the start, and occurs after commissioning, after replacing the reagent bottles or after certain diagnostic cases.