HP 8163A Series Programming Manual page 77

Lightwave multimeter, lightwave measurement system & lightwave multichannel system, mainframes
Hide thumbs Also See for 8163A Series:
Table of Contents

Advertisement

Measurement Functions – The SENSe Subsystem
command:
syntax:
description:
parameters:
response:
example:
affects:
dual sensors:
command:
syntax:
description:
parameters:
N O T E
N O T E
response:
example:
affects:
dual sensors:
command:
syntax:
description:
parameters:
response:
example:
affects:
dual sensors:
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User's Guide, E1299
:SENSe[n][:CHANnel[m]]:FUNCtion:RESult?
:SENSe[n][:CHANnel[m]]:FUNCtion:RESult?
Returns the data array of the last data acquisition function.
none
The last data acquisition function's data array as a binary block, one measurement value
is a 4-byte-long float in Intel byte order.
→ returns a data array
sens1:func:res?
All power meters and return loss modules
Master and slave channels are independent.
:SENSe[n][:CHANnel[m]]:FUNCtion:STATe
:SENSe[n][:CHANnel[m]]:FUNCtion:STATe<wsp>
LOGGing|STABility|MINMax,STOP|STARt
Enables/Disables the logging, MinMax, or stability data acquisition function mode.
LOGGing:
STABility:
MINMax:
STOP:
STARt:
When you enable a logging data acquisition function for a HP 8163A Series Power Meter
with averaging time of less than 100 ms with input hardware triggering disabled, all GPIB
commands will be ignored for the duration of the function.
See ":SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:LOGGing" on page 74 for more
information on the logging data acquisition function.
Stop any function before you try to set up a new function. Some parameters cannot be set
until you stop the function.
none
sens1:func:stat logg,star
All power meters and return loss modules
Can only be sent to master channel, slave channel is also affected.
:SENSe[n][:CHANnel[m]]:FUNCtion:STATe?
:SENSe[n][:CHANnel[m]]:FUNCtion:STATe?
Returns the function mode and the status of the data acquisition function.
none
NONE
LOGGING_STABILITY
MINMAX
PROGRESS
COMPLETE
sens1:func:stat?
All power meters and return loss modules
Can only be sent to master channel, slave channel parameters are identical.
Measurement Operations & Settings
Logging data acquisition function
Stability data acquisition function
MinMax data acquisition function
Stop data acquisition function
Start data acquisition function
No function mode selected
Logging or stability data acquisition function
MinMax data acquisition function
Data acquisition function is in progress
Data acquisition function is complete
,
LOGGING_STABILITY
COMPLETE
<END>
77

Advertisement

Table of Contents
loading

This manual is also suitable for:

8164a series8166a series

Table of Contents