Triggering - The TRIGger Subsystem
command:
syntax:
description:
parameters:
response:
example:
affects:
dual sensors:
command:
syntax:
description:
N O T E
parameters:
N O T E
response:
example:
affects:
dual sensors:
command:
syntax:
description:
parameters:
response:
example:
affects:
dual sensors:
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User's Guide, E1299
:TRIGger[n][:CHANnel[m]]:INPut?
:TRIGger[n][:CHANnel[m]]:INPut?
Returns the incoming trigger response.
none
IGNore:
Ignore incoming trigger.
SMEasure:
Start a single measurement. If a measurement function is active, see
":SENSe[n][:CHANnel[m]]:FUNCtion:STATe" on page 77, one sam-
ple is performed and the result is stored in the data array, see
":SENSe[n][:CHANnel[m]]:FUNCtion:RESult?" on page 77.
CMEasure:
Start a complete measurement. If a measurement function is active, see
":SENSe[n][:CHANnel[m]]:FUNCtion:STATe" on page 77, a complete
measurement function is performed.
NEXTstep:
Perform next step of a stepped sweep.
SWStart:
Start a sweep.
trig1:inp?
ign<END>
All tunable laser modules and power meters
Can only be sent to master channel, slave channel parameters are identical.
:TRIGger[n][:CHANnel[m]]:INPut:REARm
:TRIGger[n][:CHANnel[m]]:INPut:REARm<wsp>OFF|ON|0|1
Sets the arming response of a channel to an incoming trigger.
See Table 7, for information on how this command affects triggering power
measurements.
A boolean value:
OFF or 0: trigger rearming disabled
ON or 1: trigger rearming enabled (default)
If you return to Local control, all modules return to the default setting.
none
trig1:inp:rearm 0
All HP 8163A Series power meter modules
Can only be sent to master channel, slave channel is also affected.
:TRIGger[n][:CHANnel[m]]:INPut:REARm?
:TRIGger[n][:CHANnel[m]]:INPut:REARm?
Returns the arming response of a channel to an incoming trigger.
none
A boolean value:
0: trigger rearming disabled
1: trigger rearming enabled (default)
trig1:inp:rearm?
All HP 8163A Series power meter modules
Can only be sent to master channel, slave channel parameters are identical.
Measurement Operations & Settings