Measurement Operations & Settings
Hardware
Trigger Rearming
Trigger
trig:inp
trig:inp:rearm
IGNore
-
SMEasure
ON
CMEasure
ON
SMEasure
OFF
CMEasure
OFF
108
Triggering - The TRIGger
Subsystem
The TRIGger Subsystem allows you to configure how the instrument reacts to
incoming or outgoing triggers.
Software Triggering
init:imm
One power mea-
Automatically performs power
surement is per-
measurements.
formed.
Every hardware trigger starts a new power measure-
ment.
The first hardware trigger starts a new power mea-
surement. Further hardware triggers are ignored until
you send trig:inp:rearm again.
Table 7
Triggering and Power Measurements
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User's Guide, E1299

Triggering - The TRIGger Subsystem

Data Acquisition Functions
init:cont
MINMax
sens:func:stat
LOGGing|STABility
Automatically performs power
measurements until the func-
tion is finished.
Every hardware trigger starts a
new power measurement until
the function is finished.
The first hardware trigger starts
the function. Subsequent pow-
er measurements are automati-
cally performed until the
function is finished.
Every hardware trigger starts a
new power measurement until
the function is finished.
The first hardware trigger starts
the function. Subsequent pow-
er measurements are automati-
cally performed until the
function is finished.