HP 8163A Series Programming Manual page 113

Lightwave multimeter, lightwave measurement system & lightwave multichannel system, mainframes
Hide thumbs Also See for 8163A Series:
Table of Contents

Advertisement

Triggering - The TRIGger Subsystem
command:
syntax:
description:
parameters:
response:
example:
affects:
dual sensors:
command:
syntax:
description:
parameters:
response:
example:
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User's Guide, E1299
:TRIGger[n][:CHANnel[m]]:OUTPut:REARm?
:TRIGger[n][:CHANnel[m]]:OUTPut:REARm?
Returns the arming response of a channel to an outgoing trigger.
none
A boolean value:
0: trigger rearming disabled (default)
1: trigger rearming enabled.
trig1:outp:rearm?
All HP 8163A series power meters
Can only be sent to master channel, slave channel parameters are identical.
:TRIGger:CONFiguration
:TRIGger:CONFiguration<wsp><triggering mode>
Sets the hardware trigger configuration with regard to Output and Input Trigger Connec-
tors.
0 or DISabled:
Trigger connectors are disabled.
1 or DEFault:
The Input Trigger Connector is activated, the incoming trigger re-
sponse for each slot ":TRIGger[n][:CHANnel[m]]:INPut" on
page 110 determines how each slot responds to an incoming trigger,
all slot events (see ":TRIGger[n][:CHANnel[m]]:OUTPut" on
page 112) can trigger the Output Trigger Connector.
2 or PASSthrough:
The same as DEFault but a trigger at the Input Trigger Connector
generates a trigger at the Output Trigger Connector automatically.
3 or LOOPback:
The same as DEFault but a trigger at the Output Trigger Connector
generates a trigger at the Input Trigger Connector automatically.
none
trig:conf dis
Measurement Operations & Settings