Flash
Fault description
code
5
Left magnetic
brake fault
1602969-F
Possible action
•
Check cables and
connectors.
•
Check left magnetic
brake is engaged.
•
Refer to the chapter
"Pushing the mobility
device in freewheel
mode" in the user
manual of your
wheelchair.
•
Contact your provider.
Flash
Fault description
code
6
Right magnetic
brake fault
7
Module fault
(other than
remote module)
1 Configuration of the motors depending on the
wheelchair model
6.2 OON ("Out Of Neutral")
OON ("Out Of Neutral") is a safety feature that prevents
accidental driving or seating movements, when:
•
the system is powering up,
•
after a function change or
Troubleshooting
Possible action
•
Check cables and
connectors.
•
Check right magnetic
brake is engaged.
•
Refer to the chapter
"Pushing the mobility
device in freewheel
mode" in the user
manual of your
wheelchair.
•
Contact your provider.
•
Check cables and
connectors.
•
Check modules.
•
Recharge batteries.
•
If the chair was stalled,
reverse away or remove
obstacle.
•
Contact your provider.
33