How to deal with solvents

• Use clean bottles only.
• Exchange water- based solvents daily.
• Select solvent volume to be used up within 1 – 2 days.
• Use only HPLC- grade solvents and water filtered through 0.2 µm filters.
• Label bottles correctly with bottle content, and filling date / expiry
• Use solvent inlet filters.
• Reduce risk of algae growth: use brown bottles for aqueous solvents,

Select channels for Multi-Channel Gradient Valve (MCGV)

• Use lower channels (A and/or D) for buffer solutions.
• Regularly flush all MCGV channels with water to remove possible salt
• Check compatibility of buffers and organic solvents to avoid
Agilent 1290 Infinity II Flexible Pump User Manual
date.
avoid direct sunlight.
deposits.
precipitation.