GE Medical Systems
REV1
TABLE 3–2

ERROR MESSAGES DISPLAYED ON READOUT AND ON CONTROL CONSOLE

Message
Message on
No.
Column Readout
I 07
None
I 12
None
I 20
None
I 21
None
I 22
None
I 23
None
I 24
None
I 27
None
I 28
None
Message on
Control Console
SELECTED FILTER NOT
ALLOWED: xx
MAXIMUM mAs = xxx
AEC FAILURE
HEATING FAILURE
TRACK 1
HEATING FAILURE
TRACK 2
CONSOLE
COMMUNICATION FAIL
LINE POWER
INTERRUPTION
GENERATOR CPU
BATTERY ERROR
MAX WAIT FOR SAME
EXPOSURE: XX min
Explanation/meaning
of message
Filter selected by operator
not allowed.
Maximum available mAs are
equal to xxx.
AEC is faulty.
Molybdenum focal track is
out of use.
Rhodium focal track is out of
use.
Generator-Control Console
connection is faulty.
Micro power cut occurred on
line supply.
CPU backup battery is
discharged.
It is necessary to wait the
displayed time to be able to
repeat the same exposure.
3-6
SENOGRAPHE DMR+
om 5128689-2-100
Corrective Action
- Information message only,
because Senographe
automatically selects correct
filter.
- Information message only.
(If mAs exceeded, cell aborts
exposure at 2 or 3 mAs if entire
exposure cannot be made.)
- Switch OFF, and switch ON.
If fault continues, note ALL
messages and contact
GEMS Field Service.
- You can still work in manual
mode.
- Switch OFF, and switch ON.
- If fault continues, note ALL
messages and contact
GEMS Field Service.
- You can continue using rhodium
focal track.
- Switch OFF, and switch ON.
- If fault continues, note ALL
messages and contact
GEMS Field Service.
- You can continue using
molybdenum focal track.
- Switch OFF, and switch ON.
- If fault continues, note ALL
messages and contact
GEMS Field Service.
- Press RESET to suppress
message.
- Contact GEMS Field Service.
- DO NOT REMOVE POWER
FROM EQUIPMENT.
Wait the required amount of
time to repeat exposure. If an
exposure is attempted before
this time, an abort MAY occur
(we have built in a safety margin
to ensure proper operation).
It also means that you can
ignore the message if a
different, less dense, breast is to
be imaged next.