SECTION 2
TECHNICAL SPECIFICATIONS
2–1

Electrical Specification

2–1–1
Line voltage specifications
Line frequency specifications
2–1–2
kVA load characteristics
2–1–3
Input impedance
Note:
2–1–4
Generator Output (excluding tube)
2–1–5
Duty Cycle
The Senographe DMR is a class I type B equipment.
D
Single–phase input voltage:
* 200/208/220/240/380/415 V (± 10 %)
D
50 or 60 Hz (1 Hz)
D
8.5 kVA for exposures up to 4.5 s.
7.5 kVA during 10 s
D
Power factor: 0.61.
D
The maximum line current corresponds to the use of the technique factors 30 kV, Mo
track, large focal spot, and at least 100 mAs.
Line impedance at maximum workload 3.5%.
Line impedances under single–phase installation:
* R
= total line resistance (two-wire) of the circuit,
I
* R
380 = line resistance equivalent to R reduced to 380 V (input voltage),
I
* U = local input voltage
R
* typical value: R
380 v 1.2 W at P max = 8.5 kVA.
I
The optional STEREOTIX 2 is supplied through the SENOGRAPHE DMR gantry
(Nominal voltage = 27 V).
D
22 thru 49 kV,
D
20 thru 130 mA.
The generator is always limited by the tube:
4-6
2
380 = R
x (380/U)
I
I
SENOGRAPHE DMR+
om 5128689-2-100