4
AOP, AEC, and Manual Mode selection
The key selects the AOP, AEC and manual modes successively.
D
The AOP mode is broken down into three positions, allowing three priorities:
CNT:
contrast priority.
STD:
standard (dose/contrast compromise).
DOSE: dose reduction priority.
Pressing the key selects the Contrast, Standard, and Dose priorities. At the
start of compression, the three priorities are displayed on the kV readout.
When the exposure is complete, the initial letter of the code selected is
displayed on the kV readout.
The focal track and filter are selected automatically (pressing the key for each has
no effect). The following can be selected via their keys: screen-film combination,
film density control.
BREAST COMPRESSION IS ESSENTIAL IN AOP MODE.
D
The AEC Mode is used to select the kV values manually, and mAs values
automatically.
The following selections are available via their keys: focal track, filter, focal
spot, and film density control.
D
Manual mode is used to select kV and mAs values.
The Film Density Control key and the Screen Film Combination key have
no effect in manual mode.
1-31