GE SENOGRAPHE DMR+ Technical Publications page 69

Hide thumbs Also See for SENOGRAPHE DMR+:
Table of Contents

Advertisement

GE Medical Systems
REV1
Message
Message on
No.
Column Readout
I 86
DECOMPRESS
I 90
None
I 98
COLUMN FAULT
I 99
COLUMN FAULT
E 01
None
E 02
None
Message on
Control Console
DECOMPRESSION
AUTODECOMPRESSION
IS OFF
COLUMN DOWN
IMPOSSIBLE
None
GENERATOR FAILURE
FOCUS BIAS FAILURE
3-9
Explanation/meaning
of message
Compression paddle hit an
obstacle during
up-movement at start-up.
Compression not possible
The automatic
decompression mode has
been inhibited by the user.
Elevator up-movement is
not possible.
Elevator down-movement is
not possible.
Failure in HV section of
generator.
Failure in tube focus bias
circuits.
SENOGRAPHE DMR+
om 5128689-2-100
Corrective Action
1. Raise paddle to park using
decompression pedal.
2. Then compression is possible.
- If autodecompress mode is
wanted, switch to the opposite
mode by changing the
parameter in the Setup/Medical
menu. Otherwise nothing is
required and this information is
normally displayed.
1.Check if any movement
switches (up/down/rotation/...)
were pressed during power
ON. If yes, release the
necessary switches and restart
the system.
2.Check if obstacle preventing
up-movement.
3.Command down-movement.
4.Command up-movement.
5.If fault continues, note ALL
messages and contact
GEMS Field Service.
1.Check if any movement
switches (up/down/rotation/...)
were pressed during power
ON. If yes, release the
necessary switches and restart
the system.
2.Check if obstacle preventing
down-movement.
3.Command up-movement.
4.Command down-movement.
5.If fault continues, note ALL
messages and contact
GEMS Field Service.
- Switch OFF, and switch ON.
If fault continues, note ALL
messages and contact
GEMS Field Service.
- Switch OFF, and switch ON.
If fault continues, note ALL
messages and contact
GEMS Field Service.