REV1
4–7

Exposure Interlock Disable

Note:
4–7–1
Access to Programming Mode
4–7–2
Selecting a Function
The Senographe DMR includes a cassette exposure interlock which prevents an exposure
from being made if there is no cassette in the bucky or if the cassette used for the previous
exposure has not been removed.
Trying to make an exposure without a cassette in the bucky cassette–holder or without
having changed the cassette between two exposures will result in a beep sound and a
message will be displayed on the column and the control console and the exposure will be
inhibited.
However, it is possible, for special purposes such as physicist testing, to disable this
interlock.
The Senographe DMR AUTOMATICALLY enables exposure interlock when
powered ON.
D
To access the programming mode, press
D
The APPLICATION Menu, focal spot, focal track, filter, etc., disappears.
Press, and key lights up
To select a function, press the key located under the symbol for that function.
MEDICAL LANGUAGE
Press for rest of program
3-27
SENOGRAPHE DMR+
om 5128689-2-100
.
APPLICATION
SETUP
APPLICATION
SETUP