GE SENOGRAPHE DMR+ Technical Publications page 73

Hide thumbs Also See for SENOGRAPHE DMR+:
Table of Contents

Advertisement

GE Medical Systems
REV1
Message
Message on
No.
Column Readout
S 08
None
S 09
None
S 10
None
S 11
None
S 13
None
S 14
None
S 16
None
S 25
None
S 26
None
S 31
None
Message on
Control Console
REQUIRED POWER IS
TOO HIGH
NO COMPRESSION
DETECTED
GENERATOR COOLING
xxxxx mn
TUBE COOLING
xxxxx mn
BACKUP mAs < 50
EXPOSURE INHIBITED
MAXIMUM mAs
EXCEEDED
SELECTED
FILM/SCREEN IS
UNCALIBRATED
TUBE HOUSING OVER
TEMPERATURE
EXPOSURE TOO LONG
UNEXPOSED CASSETTE
MISSING
3-13
Explanation/meaning
of message
Power requested in Manual
Mode is too high.
Breast is not compressed.
Generator requires xxxxx
minutes to cool.
Tube requires xxxxx minutes
to cool.
Maximum mAs < 50.
Exposure inhibited.
Maximum mAs set at
installation are exceeded.
(This message appears only
if a value below 600 mAs
was calibrated at
installation).
The selected screen/film pair
is not calibrated.
The temperature of the tube
housing is too high
(> 65°C).
The exposure is too long.
This message should appear
only in manual mode.
There is no cassette, or the
cassette has not been
changed since last exposure
SENOGRAPHE DMR+
om 5128689-2-100
Corrective Action
- Lower kV or mAs values.
- Breast compression is
mandatory.
- Wait for cooling to terminate
(or lower mAs values in
Manual Mode).
- Wait for cooling to terminate
(or lower mAs values in
Manual Mode).
- Wait for cooling to terminate
(or lower kV values in AEC
Mode).
- Lower mAs values.
- Press the screen/film selection
button to select a calibrated
pair.
- If the fault continues, it
indicates that no pair has
been calibrated.
- If fault continues, note ALL
messages and contact
GEMS Field Service.
- Note ALL messages and
contact GEMS Field Service.
- Wait for the tube housing
temperature to fall.
- Increase the kV value or reduce
the mAs value.
- If a previously exposed
cassette is still in the bucky,
remove it and insert an
unexposed one.
- If there is no cassette in the
bucky, insert an unexposed
one.
- If the message is still displayed,
check that the bucky is
correctly inserted (see
chapter 1, section 2-1-3).