GE SENOGRAPHE DMR+ Technical Publications page 45

Hide thumbs Also See for SENOGRAPHE DMR+:
Table of Contents

Advertisement

GE Medical Systems
REV1
Note:
Note:
8
Film Density Control (FDC) decrements
FDC
In AOP and AEC modes, press key to reduce density correction by one step.
9
Film Density Control (FDC) increments
FDC+
In AOP and AEC modes, press key to increase density correction by one step.
FDC+, FDC-: pressing these keys in manual mode has no effect.
Correction is available in 11 steps (-5 thru +5). Density variation is 0.2 OD (optical
density) by default for each step and may be programmed at installation.
Screen film Combination (SFC) selection:
10
D
In AOP and AEC modes, press key to select screen-film combination speed.
D
The cycle turns on the number of SFCs calibrated at installation.
D
A maximum of five SFCs can be calibrated.
D
The name of the SFC (specified at installation) appears on the readout.
Pressing this key in manual mode has no effect.