Endress+Hauser HART Proline Promag 51 Operating Instructions Manual
Endress+Hauser HART Proline Promag 51 Operating Instructions Manual

Endress+Hauser HART Proline Promag 51 Operating Instructions Manual

Electromagnetic flow measuring system
Table of Contents

Advertisement

Quick Links

Operating Instructions
Proline Promag 51
HART
Electromagnetic flow measuring system
6
BA00137D/06/EN/03.10
71111278
Valid as of version
V 2.03.XX (device software)

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser HART Proline Promag 51

  • Page 1 Operating Instructions Proline Promag 51 HART Electromagnetic flow measuring system BA00137D/06/EN/03.10 71111278 Valid as of version V 2.03.XX (device software)
  • Page 3: Table Of Contents

    Spare parts ....... 73 Post-connection check ..... . 40 Endress+Hauser...
  • Page 4 10.1.14 Documentation ....91 Index ......92 Endress+Hauser...
  • Page 5: Safety Instructions

    • The device must be operated by persons authorized and trained by the facility's owner-operator. Strict compliance with the instructions in the Operating Manual is mandatory. • Endress+Hauser is willing to assist in clarifying the chemical resistance properties of parts wetted by special fluids, including fluids used for cleaning. However, small changes in temperature, concentration or the degree of contamination in the process can result in changes of the chemical resistance properties.
  • Page 6: Operational Safety

    If the temperature of the fluid is high, ensure staff are protected against burns and scalds. • The manufacturer reserves the right to modify technical data without prior notice. Your Endress+Hauser distributor will supply you with current information and updates to these Operating Instructions.
  • Page 7: Identification

    PULSE-OUT: with pulse/status output STATUS-IN: with status input (power supply) STATUS-OUT: with status output (switch output) Reserved for information on special products Reference to Operating Instructions Reserved for additional information on device version (approvals, certificates) Ambient temperature range Degree of protection Endress+Hauser...
  • Page 8: Nameplate Of The Sensor

    Reference to Operating Instructions Reserved for additional information on device version (approvals, certificates) See "Maximum measured error"  ä 84 Additional information (examples): – EPD/MSU: with Empty Pipe Detection electrode – R/B: with reference electrode Degree of protection Flow direction Endress+Hauser...
  • Page 9: Nameplate For Connections

    EMC requirements of IEC/EN 61326. The measuring system described in these Operating Instructions is therefore in conformity with the statutory requirements of the EC Directives. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.
  • Page 10: Registered Trademarks

    Registered trademark of Ladish & Co., Inc., Kenosha, USA ® HART Registered trademark of HART Communication Foundation, Austin, USA ® ® ® HistoROM™, S-DAT®, Field Xpert™, FieldCare , Fieldcheck , Applicator Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH Endress+Hauser...
  • Page 11: Installation

    Risk of injury if the measuring device slips. The center of gravity of the assembled measuring device might be higher than the points around which the slings are slung. At all times, therefore, make sure that the device does not unexpectedly turn around its axis or slip. A0004294 Transporting transmitters with DN 300 Fig. 5: Endress+Hauser...
  • Page 12: Storage

    • The measuring device must be protected against direct sunlight during storage in order to avoid unacceptably high surface temperatures. • Choose a storage location where moisture does not collect in the measuring device. This will help prevent fungus and bacteria infestation which can damage the liner. Endress+Hauser...
  • Page 13: Installation Conditions

    It might be necessary to install pulse dampers in systems incorporating reciprocating, diaphragm or peristaltic pumps. Information on the measuring system's resistance to vibration and shock can be found on  ä 84. A0003203 Fig. 8: Installation of pumps Endress+Hauser...
  • Page 14 This measure also prevents the system losing prime, which could cause air inclusions. Information on the lining's resistance to partial vacuum can be found on  ä 85. A0008157 Fig. 10: Measures for installation in a down pipe Vent valve Pipe siphon Length of down pipe Endress+Hauser...
  • Page 15: Orientation

    Empty Pipe Detection will respond if the measuring tube is only partially filled or empty. A0003207 Fig. 12: Horizontal orientation EPD electrode for the detection of empty pipes Measurement electrodes for the signal acquisition Reference electrode for the potential equalization Endress+Hauser...
  • Page 16: Vibrations

    Caution! It is advisable to install sensor and transmitter separately if vibration is excessively severe. Information on resistance to vibration and shock  ä 84. A0003208 Fig. 14: Measures to prevent device vibrations (L > 10 m (32.8 ft)) Endress+Hauser...
  • Page 17: Foundations, Supports

    From the nomogram read off the pressure loss as a function of flow velocity (downstream from the reduction) and the d/D ratio. [mbar] 8 m/s 7 m/s 6 m/s 5 m/s 4 m/s max. 8° 3 m/s 2 m/s 1 m/s d / D A0003213 Fig. 16: Pressure loss due to adapters Endress+Hauser...
  • Page 18: Flow Characteristics

    Nominal Q3/Q1 Diameter DN [mm] [m³/h] [m³/h] [m³/h] [m³/h] [m³/h] 0.04 0.06 0.10 0.16 0.16 0.25 0.25 0.40 0.39 0.63 0.63 1000 1250 1600 2000 2500 3125 2500 3125 4000 5000 6300 7875 6300 7875 6300 7875 6300 7875 Endress+Hauser...
  • Page 19: Length Of Connecting Cable

    • The maximum length of the connecting cable is 10 m (32.8 ft) when Empty Pipe Detection (EPD  ä 59) is switched on. [µS/cm] [ft] A0010734 Fig. 17: Permissible cable length for the remote version Gray shaded area = permissible range = length of connecting cable in [m] Fluid conductivity in [μS/cm] Endress+Hauser...
  • Page 20: Installation Instructions

    • If necessary, a special ground cable for potential equalization can be ordered as an accessory ( ä 62). • Information on potential equalization and detailed installation instructions for using ground cables can be found on  ä 37. Endress+Hauser...
  • Page 21 20 × M 33 PN 25 20 × M 36 PN 6 24 × M 24 PN 10 24 × M 27 PN 16 24 × M 33 PN 25 24 × M 39 PN 6 24 × M 27 Endress+Hauser...
  • Page 22: Installing The Promag P Sensor

    Ground cable • If necessary, special ground cables for potential equalization can be ordered as an accessory ( ä 62). • Information on potential equalization and detailed installation instructions for using ground cables can be found on  ä 37. Endress+Hauser...
  • Page 23 PN 25 20 × M 33 – PN 10 20 × M 27 – 600 * PN 16 20 × M 33 – PN 25 20 × M 36 – * Designed acc. to EN 1092-1 (not to DIN 2501) Endress+Hauser...
  • Page 24: Turning The Transmitter Housing

    Carefully lift the transmitter housing as far as it will go. Turn the transmitter housing to the desired position (max. 2 × 90° in either direction). Put the housing back on. Retighten the two securing screws. a0004303 Fig. 21: Turning the transmitter housing (stainless steel field housing) Endress+Hauser...
  • Page 25: Turning The Local Display

    Rotate the display to the desired position (max. 4 × 45° in each direction), and place it back onto the electronics compartment cover plate. Screw the cover of the electronics compartment firmly onto the transmitter housing. 4 x 45° a0003236 Fig. 22: Turning the local display (field housing) Endress+Hauser...
  • Page 26: Installing The Wall-Mount

    – Screw head: max. Ø 10.5 mm (0.41") Secure the transmitter housing to the wall as indicated. Screw the cover of the connection compartment (a) firmly onto the housing. 35 (1.38) 90 (3.54) 192 (7.56) mm (inch) a0001130 Fig. 23: Mounted directly on the wall Endress+Hauser...
  • Page 27 If the device is mounted to a warm pipe, make certain that the housing temperature does not exceed +55 °C (+131 °F), which is the maximum permissible temperature. Ø 20…70 (Ø 0.79…2.75) 155 ( ~ 6.1) mm (inch) a0001132 Fig. 25: Pipe mounting (wall-mount housing) Endress+Hauser...
  • Page 28: Post-Installation Check

    Are the inlet and outlet runs to respected? Outlet run  2 × DN Is the measuring device protected against moisture and direct sunlight? Is the sensor adequately protected against vibration (attachment, support)? Acceleration up to 2 g by analogy with IEC 600 68-2-8 Endress+Hauser...
  • Page 29: Wiring

    When connecting Ex-certified devices, see the notes and diagrams in the Ex-specific supplement to these Operating Instructions. Please do not hesitate to contact your Endress+Hauser representative if you have any questions. Note! The measuring device does not have an internal circuit breaker. For this reason, assign the measuring device a switch or power-circuit breaker which can be used to disconnect the power supply line from the power grid.
  • Page 30 Wall-mount housing connection compartment Cover of the sensor connection housing Signal cable Coil current cable n.c. Not connected, insulated cable shields Wire colors/Terminal No.: 5/6 = braun, 7/8 = white, 4 = green, 37/36 = yellow, 41 = 1, 42 = 2 Endress+Hauser...
  • Page 31 A0002688-ae SENSOR Signal cable Coil current cable 20 (0.79)* 170 (6.69)* mm (inch) 20 (0.79)* 160 (6.30)* mm (inch) 80 (3.15) 70 (2.76) 17 (0.67) 50 (1.97) 50 (1.97) 8 (0.31) 10 (0.39) 8 (0.31) ³1 (0.04) A0002646-AE A0002650-ae Endress+Hauser...
  • Page 32: Cable Specifications

    Outer jacket Reinforced connecting cables As an option, Endress+Hauser can also deliver reinforced connecting cables with an additional, reinforcing metal braid. Reinforced connecting cables should be used when laying cables under ground, in situations where there is the danger of cable destruction by rodents and if using a measuring device with IP 68 degree of protection.
  • Page 33: Connecting The Measuring Unit

    Terminal No. 2: N for AC, L- for DC Signal cable: Terminals Nos. 20–27  ä 34 Ground terminal for protective conductor Ground terminal for signal cable shield Service adapter for connecting service interface FXA193 (Fieldcheck, FieldCare) Cover of the connection compartment Securing clamp Endress+Hauser...
  • Page 34: Terminal Assignment

    24 (+) / 25 (–) 26 (+) / 27 (–) 51***-***********D Current output Status input Status output Frequency output HART Pulse input Current output 51***-***********P Status input Status output custody transfer HART Note! Functional values of the inputs and outputs  ä 80 Endress+Hauser...
  • Page 35: Connecting A "Reset Key

    A = Wiring version with external voltage supply (3 to 30 V) B = Wiring version with the current output as voltage supply (4 to 20 mA) Terminal No. 20 / 21 = Status input Terminal No. 26 / 27 = Current output Endress+Hauser...
  • Page 36: Hart Connection

    In order to connect a PC with operating software (e.g. "FieldCare"), a HART modem (e.g. "Commubox FXA195") is needed. ³ –27 a0004592 Fig. 32: Electrical connection of a PC with operating software PC with operating software Auxiliary energy Shielding Other devices or PLC with passive input HART modem, e.g. Commubox FXA195 Endress+Hauser...
  • Page 37: Potential Equalization

    • DN  350: The ground cable is mounted directly on the metal transport bracket. Note! The ground cable for flange-to-flange connections can be ordered separately as an accessory from Endress+Hauser. A0011893 Fig. 34: Via the ground terminal of the...
  • Page 38 • There should be no electrically conductive connection A0011896 between the pipe and the device. Fig. 36: Potential equalization and cathodic • The mounting material must withstand the applicable protection (ground cable: copper wire torques. at least 6 mm²/0.0093 in²) Power supply isolation transformer Electrically isolated Endress+Hauser...
  • Page 39: Degree Of Protection

    Caution! Do not loosen the threaded fasteners of the sensor housing, as otherwise the degree of protection guaranteed by Endress+Hauser no longer applies. Note! The Promag W and Promag P sensors can be supplied with IP 68 rating (permanent immersion in water to a depth of 3 meters (10ft)).
  • Page 40: Post-Connection Check

    Are all screw terminals firmly tightened?  ä 37 Have the measures for grounding/potential equalization been correctly implemented?  ä 39 Are all cable entries installed, firmly tightened and correctly sealed? Cables looped as "water traps"? Are all housing covers installed and firmly tightened? Endress+Hauser...
  • Page 41: Operation

    – Press and hold down +/- keys for longer than 3 seconds  Return directly to HOME position – Cancel data entry Enter key – HOME position  Entry into the function matrix – Save the numerical values you input or settings you change Endress+Hauser...
  • Page 42: Brief Guide To The Function Matrix

    – Press and hold down Esc key (X) for longer than 3 seconds  HOME position – Repeatedly press Esc key (X)  return step by step to HOME position > 3 s – – – – A0001142 Fig. 39: Selecting functions and configuring parameters (function matrix) Endress+Hauser...
  • Page 43 • If the power supply fails, all preset and parameterized values remain safely stored in the EEPROM. " Caution! All functions are described in detail, including the function matrix itself, in the "Description of Device Functions" manual, which is a separate part of these Operating Instructions. Endress+Hauser...
  • Page 44: Displaying Error Messages

    The response of the individual outputs (failsafe mode) can be defined by means of the "FAILSAFE MODE" function in the function matrix ( "Description of Device Functions" manual). Note! For safety reasons, error messages should be output via the status output. Endress+Hauser...
  • Page 45: Communication

    Operating program "FieldCare" FieldCare is Endress+Hauser’s FDT-based plant asset management tool and allows the configuration and diagnosis of intelligent field devices. By using status information, you also have a simple but effective tool for monitoring devices. The Proline flow measuring devices are accessed via a service interface or via the service interface FXA193.
  • Page 46: Current Device Description Files

    The "Fieldcheck" tester/simulator is used for testing flowmeters in the field. When used in conjunction with the "FieldCare" software package, test results can be imported into a database, printed and used for official certification. Contact your Endress+Hauser representative for more information.
  • Page 47: Universal And Common Practice Hart

    • Third process variable = OFF (not assigned) • Fourth process variable = OFF (not assigned) Note! • Manufacturer-specific units are represented using the HART unit code "240". • You can change the assignment of device variables to process variables using Command 51. Endress+Hauser...
  • Page 48: Commands

    – Byte 16: OEM dealer ID, 17 = E+H Factory setting: Primary process variable = Volume flow Note! • Manufacturer-specific units are represented using the HART unit code "240". • You can change the assignment of device variables to process variables using Command 51. Endress+Hauser...
  • Page 49 Bytes 0-3: Output current in mA Bytes 0-3: Output current in mA Factory setting: Primary process variable = Volume flow Note! You can set the assignment of device variables to process variables using Command 51. Perform master reset none none Access = write Endress+Hauser...
  • Page 50 As a response, the current number of the preambles is response message are inserted in the response messages: displayed in the response message: Byte 0: Number of Access = write Byte 0: Number of preambles (4 to 20) preambles Endress+Hauser...
  • Page 51: Device Status And Error Messages

    60 seconds. Pulse buffer: The buffering of the flow components (measuring mode with pulsating flow) could not be calculated and output within 60 seconds. Current output: Flow is out of range. Frequency output: Flow is out of range. Endress+Hauser...
  • Page 52 Uploading/downloading device files. Currently no other commands are possible. Positive zero return active Simulation current output active Simulation frequency output active Simulation pulse output active Simulation status output active Simulation of the status input active Simulation of response to error (outputs) active Simulation of volume flow active Endress+Hauser...
  • Page 53: Commissioning

    These functions are indicated by a keyhole symbol (`) in the separate device functions manual. Screw on the cover of the electronics compartment. Lead-seal the device  å 41. The device is in custody transfer mode. The flow measurement values can now be used in commercial transactions. Endress+Hauser...
  • Page 54: Deactivating Custody Transfer Mode

    During startup, the message "CUSTODY TRANSFER NO" appears on the display. 12. The device is now operational and is not in custody transfer mode. Note! To set the device back to the custody transfer mode:  ä 53, Setting-up custody transfer mode. Endress+Hauser...
  • Page 55: Special Features Of Operation Subject To Metrological Control/Custody Transfer Measurement

    Flow ranges and error limits for cold water 1 = Lower range (error limit: ±5%) 2 = Upper range (error limit: ±2%) Minimum flowrate Q Lowest flowrate at which the water meter is required to operate within the maximum permissible error. Endress+Hauser...
  • Page 56: Function Check

    • If a device is in the custody transfer mode, the error message "POWER BRK. DWN" appears on the display each time the power supply is applied. This error message has to be reset and acknowledged via the status input. Each time an error is reset, a display test is triggered. Endress+Hauser...
  • Page 57: Quick Setup

    Width Output Output Time Signal Signal Constant Failsafe Time Failsafe Mode Constant Mode Failsafe Mode Automatic parameterization of the display Quit Quick Setup A0005413-en Fig. 43: "QUICK SETUP COMMISSIONING" menu for the rapid configuration of important device functions Endress+Hauser...
  • Page 58: Configuration

    Incorrectly set jumpers can cause overcurrents that would destroy either the measuring device or external devices connected to it. Installation of the I/O board is the reverse of the removal procedure. A0001044 Fig. 44: Configuring the current outputs with the jumpers (I/O board) Active current output (factory setting) Passive current output Endress+Hauser...
  • Page 59: Adjustment

    The adjustment values for empty pipe and full pipe are identical. In cases of this nature you must repeat empty-pipe or full-pipe adjustment! – ADJUSTMENT NOT OK Adjustment is not possible because the fluid’s conductivity is out of range. Endress+Hauser...
  • Page 60: Data Storage Device (Historom)

    Proline Promag 51 Data storage device (HistoROM) At Endress+Hauser, the term HistoROM refers to various types of data storage modules on which process and measuring device data are stored. By plugging these modules into other slots, device configurations can be duplicated onto other measuring devices to cite just one example.
  • Page 61: Maintenance

    Proline Promag 51 Maintenance Maintenance No special maintenance work is required. Exterior cleaning When cleaning the exterior of measuring devices, always use cleaning agents that do not attack the surface of the housing and the seals. Endress+Hauser...
  • Page 62: Accessories

    Accessories Proline Promag 51 Accessories Various accessories, which can be ordered separately from Endress+Hauser, are available for the transmitter and the sensor. The Endress+Hauser service organization can provide detailed information on the order codes of your choice. Measuring principle-specific accessories...
  • Page 63: Communication-Specific Accessories

    SFX100 - ******* handheld terminal measured values via the current output HART (4 to 20 mA) and FOUNDATION Fieldbus. Contact your Endress+Hauser representative for more information. Fieldgate FXA320 Gateway for remote interrogation of HART sensors and actuators FXA320 - ***** via Web browser: •...
  • Page 64: Troubleshooting

    • For certified devices subject to legal metrological control, faults can only be corrected by breaking the seal. After repair by an Endress+Hauser service technician (provisional seal), the device must be resealed by the appropriate standards authority.
  • Page 65 • Also observe the information on  ä 44! • The measuring system interprets simulations and positive zero return as system errors, but displays them as notice messages only.  Other error (without error message) Diagnosis and rectification  ä 70 Some other error has occurred. Endress+Hauser...
  • Page 66: System Error Messages

    In the event of a serious fault, a flowmeter might have to be returned to the manufacturer for repair. The procedures on  ä 6 must be carried out before you return a flowmeter to Endress+Hauser. Always enclose a duly completed "Declaration of Contamination" form. You will find a preprinted blank of the form at the back of this manual.
  • Page 67 !: # 355 to 358 Flow is out of range. 2. Increase or reduce flow, as applicable. Custody transfer mode Change the upper or lower limit setting via the HART interface or by means of the FieldCare operating software. Endress+Hauser...
  • Page 68 Simulation status input active Switch off simulation. … !: # 671 to 674 Custody transfer mode Simulation not possible S: SIM. FAILSAFE Simulation of response to error (outputs) active Switch off simulation. !: # 691 Custody transfer mode Simulation not possible Endress+Hauser...
  • Page 69: Process Error Messages

    Repeat adjustment, making sure procedure is correct  ä 59. FULL = EMPTY The EPD adjustment values for empty pipe and full pipe $: # 463 are identical, therefore incorrect. Custody transfer mode In custody transfer mode, an EPD/OED adjustment is not possible. Endress+Hauser...
  • Page 70: Process Errors Without Messages

    Remarks: You may have to change or correct certain settings in functions in the function matrix in order to rectify the fault. Note! For certified devices subject to legal metrological control, faults can only be corrected by breaking the lead seal or protection sticker. Contact your Endress+Hauser representative to repair the device.
  • Page 71: Response Of Outputs To Errors

    Output of the frequency specified in the FAILSAFE VALUE function. HOLD VALUE Measured value display on the basis of the last saved value preceding occurrence of the fault. ACTUAL VALUE Measured value display on the basis of the current flow measurement. The fault is ignored. Endress+Hauser...
  • Page 72 The totalizers continue to count the flow in accordance with the last valid flow value (before the error occurred). Status output In the event of a fault or power supply failure: No effect on status output Status output  non-conductive fixed setting in custody transfer mode Endress+Hauser...
  • Page 73: Spare Parts

    Note! You can order spare parts directly from your Endress+Hauser service organization by providing the serial number printed on the transmitter's nameplate ( ä 7). Spare parts are shipped as sets comprising the following parts: •...
  • Page 74: Removing And Installing

    • When connecting Ex-certified devices, see the notes and diagrams in the Ex-specific supplement to these Operating Instructions. " Caution! Use only original Endress+Hauser parts. Switch off power supply. Unscrew cover from the transmitter housing. Remove the local display (1) as follows: –...
  • Page 75 Field housing: removing and installing printed circuit boards Local display Latch Ribbon cable (display module) Screws of electronics compartment cover Aperture for tool, removal/installation Power supply board Amplifier board Electrode signal cable (sensor) Coil current cable (sensor) Histo-ROM / S-DAT (sensor data memory) I/O board Endress+Hauser...
  • Page 76 • When connecting Ex-certified devices, see the notes and diagrams in the Ex-specific supplement to these Operating Instructions. " Caution! Use only original Endress+Hauser parts. Switch off power supply. Remove the screws and open the hinged cover (1) of the housing. Remove screws of the electronics module (2).
  • Page 77 Housing cover Electronics module Ribbon cable (display module) Cover of electronics compartment (3 screws) Aperture for tool, removal/installation Power supply board Amplifier board Electrode signal cable (sensor) Coil current cable (sensor) Histo-ROM / S-DAT (sensor data memory) I/O board Endress+Hauser...
  • Page 78: Replacing The Device Fuse

    – Power supply 85 to 260 V AC  0.8 A slow-blow / 250 V; 5.2 × 20 mm – Ex-rated devices  see the Ex documentation. Installation is the reverse of the removal procedure. " Caution! Use only original Endress+Hauser parts. a0001148 Fig. 48: Replacing the device fuse on the power supply board Protective cap...
  • Page 79: Return

    Costs incurred for waste disposal and injury (burns, etc.) due to inadequate cleaning will be charged to the owner-operator. The following steps must be taken before returning a flow measuring device to Endress+Hauser, e.g. for repair or calibration: • Always enclose a duly completed "Declaration of contamination" form. Only then can Endress+Hauser transport, examine and repair a returned device.
  • Page 80: Technical Data

    • U = 3 to 30 V DC • Ri = 5 k • Configurable for: totalizer reset, positive zero return, error message reset In the custody transfer mode, error messages can only be reset and a display test function activated via the status input! Endress+Hauser...
  • Page 81: Output

    • Configurable for: error messages, Empty Pipe Detection (EPD), flow direction, limit values Low flow cut off • Low flow cutoff • Switch points permanently set (On at 0.02 m/s; Off at 0.04 m/s) Galvanic isolation All circuits for inputs, outputs, and power supply are galvanically isolated from each other. Endress+Hauser...
  • Page 82: Power Supply

    • Lasting min. 1 power cycle • EEPROM saves measuring system data • S-DAT: exchangeable data storage chip which stores the data of the sensor (nominal diameter, serial number, calibration factor, zero point, etc.)  ä 37 Potential equalization Endress+Hauser...
  • Page 83: Operating Conditions: Rated Operating Conditions

    • Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic regions. • If both fluid and ambient temperatures are high, install the transmitter at a remote location from the sensor. Electromagnetic ambient Class: E2 conditions Endress+Hauser...
  • Page 84 Shock and vibration resistance Acceleration up to 2 g by analogy with IEC 60068-2-6 (high-temperature version: no data available) Electromagnetic compatibility • To IEC/EN 61326 and NAMUR recommendation NE 21. (EMC) • Emission: As per industry limit values, EN 55011 Endress+Hauser...
  • Page 85: Operating Conditions: Process

     ä 18 Limiting flow Pressure loss • No pressure loss if the sensor is installed in a pipe of the same nominal diameter. • Pressure losses for configurations incorporating adapters according to DIN EN 545 (see "Adapters"  ä 17). Endress+Hauser...
  • Page 86: Mechanical Construction

    30" – – 32" Transmitter Promag (compact version): 3.4 kg (Weight data valid for standard pressure ratings and without packaging material) *Only DN sizes 80, 100, 150 to 400, 500 and 600 are available for flanges as per AS) Endress+Hauser...
  • Page 87 • Flanges EN 1092-1 (DIN2501): 1.4571/316L; RSt37-2 (S235JRG2); C22; FE 410W B (DN  300 with Al/Zn protective coating; DN  350 with protective lacquering) • Seals: Seals to DIN 1514-1 • Ground disks: 1.4435/316L, Alloy C-22, tantalum, titanum Endress+Hauser...
  • Page 88 – DN 65 PN 16 and DN 600 PN 16 only in accordance with EN 1092-1 Surface roughness All data relate to parts in contact with fluid. • Liner PFA:  0.4 μm (15 μin) • Electrodes: 0.3 to 0.5 μm (12 to 20 μin) Endress+Hauser...
  • Page 89: 10.1.10 Human Interface

    CE mark The measuring system is in conformity with the statutory requirements of the EC Directives. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. C-tick mark The measuring system meets the EMC requirements of the Australian Communications and Media Authority (ACMA).
  • Page 90 PED is required, this must be ordered explicitly. For devices with nominal diameters less than or equal to DN 25 (1"), this is neither possible nor necessary. • With the identification PED/G1/III on the sensor nameplate, Endress+Hauser confirms conformity with the "Basic safety requirements" of Appendix I of the Pressure Equipment Directive 97/23/EC.
  • Page 91: 10.1.12 Ordering Information

    (0/4 to 20 mA), pulse value, display mode or totalizer units! 10.1.13 Accessories Various accessories, which can be ordered separately from Endress+Hauser, are available for the transmitter and the sensor  ä 62. The Endress+Hauser service organization can provide detailed information on the order codes of your choice.
  • Page 92: Index

    Error message types ......44 Promag W........20 Endress+Hauser...
  • Page 93 Pressure measuring device approval ....90 Promag W ........21 Endress+Hauser...
  • Page 94 Wiring........29 Endress+Hauser...
  • Page 95 Erklärung zur Kontamination und Reinigung Please reference the Return Authorization Number (RA#), obtained from Endress+Hauser, on all paperwork and mark the RA# clearly on the outside of the box. If this procedure is not followed, it may result in the refusal of the package at our facility.
  • Page 96 www.endress.com/worldwide BA00137D/06/EN/03.10 71111278 FM+SGML6.0 ProMoDo...

This manual is also suitable for:

Proline promag 51

Table of Contents