Dos Priming - Miele PG 8583 CD Operating Instructions Manual

Laboratory glassware washer for laboratory glassware and utensils
Hide thumbs Also See for PG 8583 CD:
Table of Contents

Advertisement

Additional settings

DOS priming

86
The dispensing system for liquid process chemicals can only dis-
pense reliably if the system has been purged of air.
The DOS system must be primed only if:
- It is being used for the first time.
- The reservoir was exchanged.
- The dispensing system has been emptied completely.
Before priming, ensure that the liquid process chemical container is
sufficiently full and the siphons are securely screwed to the contain-
ers. Only one DOS system can be primed at a time.
Open the menu as follows:
 button
Additional settings
     
Dispensing system
         
             
DOS...
(name of dispensing system)
DOS priming
                 
DOS priming
Automatic priming will start when the dispensing system is selected.
Once started, the automatic priming process can no longer be can-
celed.
Select a dispensing system using the  and  arrow buttons.
Press OK to start the priming process.
Automatic priming is successfully completed when the following mes-
sage appears in the display:
DOS
Dispensing system
successfully primed

Hide quick links:

Advertisement

Table of Contents
loading