Program settings
104
The action desired at different spray arm rotational speeds is set in
each wash block.
   ...
         Select wash block, e.g.:
Rinse arm monitoring
             
Rinse arm monitoring
On -> Warning
On -> Stop
-
A program in operation is interrupted if the rotational speed devi-
ates. The interruption is shown on the display and noted in the cy-
cle report.
On -> Warning
-
The program continues to run normally if the rotational speed devi-
ates. Only a message appears on the display and the deviation is
noted in the cycle report.
-
Off
No message appears and the program continues to run normally.
Select an option using the  and  arrow buttons.
Press OK to save the setting.
Main wash 1
On -> Stop