Problem
Fault 518–521
Fault 550
Fault 555
Fault 559
Possible cause and solution
No flow was detected when dispensing from an external
container.
 Exercise caution when handling process chemicals.
For all process chemicals, the manufacturer's safety in-
structions as given on their safety data sheets must be
observed.
Check the level in the containers and replace empty
ones with filled ones.
Check the suction openings of the suction wands and
remove any deposits.
Check the hose connections on the suction wands, the
machine, and any dispenser modules.
Remove any kinks from the dispensing hoses and
check the hoses for leaks. Position the dispensing
hoses so that they cannot kink.
Vent the dispensing systems.
If you identify any leaks in the dispensing hoses or de-
fects on the suction wands, contact Miele Service.
The waterproof system has been activated. One of the wa-
ter supply hoses might have a leak.
Close the faucets.
Contact Miele Service.
Too much water has accumulated in the steam condenser.
Wash water may have been diverted behind the protective
panel of the steam condenser on the back wall of the wash
cabinet because of angled wash items or injector nozzles.
When installing angled injector nozzles and sorting wash
items, ensure that the wash water outlets point towards
the center of the wash cabinet.
Restart the machine. Excess water is pumped out auto-
matically.
There is a problem with the process documentation inter-
face. The machine has detected a module for and Ethernet
interface, but only a serial interface is activated in the con-
trols (RS232).
Deactivate the RS232 interface:
Open the menu for configuring the interface
settings
Interface
/
Wait approx. 90 seconds. The Ethernet module XKM
3000 L Med needs this time for initialization. It may be
necessary to reconfigure the interface.
Or
Replace the Ethernet module XKM 3000 L Med with a
XKM RS232 10 Med module to set up a serial interface.