Program settings
Activating con-
ductivity measur-
ing
Setting limit val-
ues
106
Conductivity measuring is activated individually for each program.
Conductivity is measured in the final rinse phase.
...
Conductivity Limit
Conductivity Limit
Water intake
-
The electrical conductivity of the water after water intake during fi-
nal rinse phase is measured.
-
Water drainage
The electrical conductivity of the water at the end of the final rinse
phase is measured.
Select an option using the and arrow buttons.
Press OK to save the setting.
Following selection a maximum limit value for electrical conductivity
must be set.
The limit values for
ally.
...
Conductivity Limit
Water intake
Set
Water intake
--
µS/cm
(5 - 200 µS/cm )
The limit values are set in increments of 1. The possible range is
shown in the bottom line of the display.
Use the arrow buttons (higher) and (lower) to set the limit
value.
Press OK to save the limit value.
Water intake
Water drainage
Water intake
and
Water drainage
Water drainage
and/or
are set individu-
Water drainage
--
µS/cm
(5 - 200 µS/cm )