Laboratory glassware washer for laboratory glassware and utensils
Hide thumbs Also See for PG 8583 CD:
Table of Contents

Advertisement

Program settings
Activating con-
ductivity measur-
ing
Setting limit val-
ues
106
Conductivity measuring is activated individually for each program.
Conductivity is measured in the final rinse phase.
   ...
Conductivity Limit
         
Conductivity Limit
Water intake
-
The electrical conductivity of the water after water intake during fi-
nal rinse phase is measured.
-
Water drainage
The electrical conductivity of the water at the end of the final rinse
phase is measured.
Select an option using the  and  arrow buttons.
Press OK to save the setting.
Following selection a maximum limit value for electrical conductivity
must be set.
The limit values for
ally.
   ...
         
Conductivity Limit
Water intake
            
Set
                 
Water intake
--
µS/cm
(5 - 200 µS/cm )
The limit values are set in increments of 1. The possible range is
shown in the bottom line of the display.
Use the arrow buttons  (higher) and  (lower) to set the limit
value.
Press OK to save the limit value.
Water intake
Water drainage
Water intake
and
Water drainage
Water drainage
and/or
are set individu-
Water drainage
--
µS/cm
(5 - 200 µS/cm )