Miele PG 8583 CD Operating Instructions Manual page 147

Laboratory glassware washer for laboratory glassware and utensils
Hide thumbs Also See for PG 8583 CD:
Table of Contents

Advertisement

Problem
Conductivity level too high:
Actual value: µS/cm Max
value: µS/cm
Conductivity level exceeds
limit
Conductivity module requires
calibration
Conductivity module commu-
nication error
Spray pressure exceeds tol-
erance
Spray pressure fluctuating too
much
Possible cause and solution
Carry-over of conductive substances during reprocessing.
Check the process.
Empty or faulty water softener or demineralization system.
Check external water softener or demineralization sys-
tems.
Reactivate the systems if necessary.
Work on the on-site water supply.
Contact a qualified plumber.
Water connections transposed.
Observe the markings on the water connections (see
"Connection to the water supply").
If the level drops below the measuring range, the conduc-
tivity cannot be determined.
Contact Miele Customer Service.
The conductivity meter must be recalibrated.
Contact Miele Customer Service.
The connection to the conductivity meter has been inter-
rupted.
Contact Miele Customer Service.
The wash pressure differs from the reference value.
Possible causes of fluctuations in the wash pressure in-
clude:
- defective water connections,
- open adapters,
- foam build-up.
Identify and resolve the cause of this.
The program is not interrupted. Nevertheless, you must
reprocess the load.
A program was interrupted because of severe fluctuations
in the wash pressure.
Possible causes of fluctuations in the wash pressure in-
clude:
- defective water connections,
- open adapters,
- foam build-up.
Identify and resolve the cause of this.
Reprocess the load again.