Repeating if limit
value is exceeded
Calibrating the
conductivity me-
ter
If the conductivity for incoming water or draining water exceeds the
limit value, the measurement can be repeated.
If the limit value for incoming water is too high, the water is pumped
out and fresh water is allowed in again. If the conductivity value at the
end of the final rinse phase exceeds the limit value for draining water,
the entire final rinse block is repeated.
If the measured value still exceeds the limit value after the repeat
measurement, the program ends and a fault message is generated.
The message can be shown on the display and also in the cycle re-
port.
   ...
         
Conductivity Limit
Water intake
            
Number of repeats
                 
The setting value is entered in increments of 1. The possible range is
shown in the bottom line of the display.
Use the arrow buttons  (higher) and  (lower) to set the number of
repetitions.
Press OK to save the setting.
At regular intervals, e.g. in the course of maintenance, the conductiv-
ity meter must be calibrated by Miele Service. Calibration outside the
maintenance cycles may also become necessary.
Conductivity module
requires calibration
If this message appears, contact Miele Service.
Water drainage
and/or
Number of repeats
1
(0 - 1)
Program settings
107