Miele PG 8583 CD Operating Instructions Manual page 140

Laboratory glassware washer for laboratory glassware and utensils
Hide thumbs Also See for PG 8583 CD:
Table of Contents

Advertisement

Frequently asked questions
Problem
Fault 438
Fault 440
Fault 460-462
Fault 492, 504
140
Possible cause and solution
The door seal sticks.
Clean the door seal.
Heavy objects in front of the machine can impede the au-
tomatic opening of the door by the Comfort lock.
Do not place (heavy) objects in front of the door of the
machine.
The Comfort door lock is blocked.
Try to open the door carefully (without using force) by
pulling on the door handle.
If the door is still blocked:
Open the door using the emergency release.
Close the door and try to open it again using the  but-
ton.
If it is still blocked:
Contact Miele Customer Service.
The float switch in the sump of the wash cabinet has not
been activated. The switch might be blocked.
Remove the filter combination.
Check the float switch to make sure it moves freely. The
float switch is located in the sump of the wash cabinet
behind the spray arm.
A program was interrupted due to the spray arm speed
dropping below the set value because:
- items are obstructing the machine or basket spray arms.
Arrange the load so that the spray arms can turn easily
and start the program again.
- wash pressure is too low due to a heavy build-up of
foam.
Follow the instructions regarding foam build-up in
"Chemical processes and technology."
A program was canceled because there is not enough
spray pressure. The filters in the wash chamber may be
blocked.
 Danger of injury from glass shards, needles etc.
which are retained in the filter.
Check and clean the filters in the wash chamber (see
"Maintenance/Cleaning the filters in the wash chamber").