Changing Water Quantity - Miele PG 8583 CD Operating Instructions Manual

Laboratory glassware washer for laboratory glassware and utensils
Hide thumbs Also See for PG 8583 CD:
Table of Contents

Advertisement

Program settings
Activating the
conductivity sen-
sor
Changing water
quantity
108
The conductivity is measured during the final rinse phase (see pro-
gram charts).
   ...
Final rinse 1
         
or
Conductivity Monitoring
             
Conductivity Monitoring
Yes
-
The conductivity is being measured.
No
-
The conductivity sensor is deactivated.
Select an option using the  and  arrow buttons.
Press OK to save the setting.
Increasing the water level is advisable if a large amount of water
clings to items due to the structure of the wash load or if a heavy
build-up of foam might occur due to the type of soiling (e.g. blood)
and the process chemicals used. The additional amount of water re-
quired depends on the type of basket or mobile unit used, the type of
soiling and the load.
If a lightly soiled load is being reprocessed which does not hold much
water, the amount of water can be reset to the factory default amount
to save water and energy.
   ...
Water volume change
         
The water quantity can be increased in 17 oz. (0.5 l) increments, or
set back to the factory default amount. The possible range is shown
in the bottom line. The setting "0 l" equates to the factory default set-
ting.
Use the arrow buttons  (higher) and  (lower) to alter the water
quantity.
Press OK to save the setting.
Final rinse 2
Yes
No
Water volume change
0.0
l
(0.0 - 8.5 l )