HOW TO:
Making Measurements
Measuring Error Analysis
M.2110 BIS (Bring Into Service)
Provides a 15 minute, 2 hour, 24 hour and 7 day PASS, FAIL, ? indication for BIS
testing as described in M.2110. The ES and SES results are compared to the S1 and
S2 thresholds and indicate PASS, FAIL or ? (uncertain). If the result is "uncertain",
the procedures laid down in M.2110 should be followed.
The S1 and S2 thresholds can be set in two different ways:
1 USER - S1 and S2 values are input by the user.
2 PATH ALLOCATION - ProBER 2 calculates the S1 and S2 values from the
user-entered Path Allocation value according to tables and procedures in M.2100.
M.2120 Circuit Maintenance
Provides a threshold report when any of the relevant thresholds are exceeded within
a 15 minute (TR1 ES and SES) or 24 hour period (TR2 ES and SES).
The TR1 and TR2 thresholds can be set in two different ways:
1 USER - TR1 ES and SES, and TR2 ES and SES values are input by the user.
2 PATH ALLOCATION - HP ProBER 2 calculates the TR1 and TR2 values from
user-entered Path Allocation and Maintenance Factor values according to
M.2120.
1 Set up the transmit and receive interfaces. See "Setting Transmit and Receive
Interfaces" on page 4-3.
2 Choose the Test Timing required (see "Setting Test Timing" on page 6-2) and
start the test by pressing
3 Choose
,
Results
MENU
M.2100
M.2110
,
6-8
.
START/STOP
,
,
Analysis
or
to obtain the Analysis Results menu, as shown.
M.2120
,
,
, then
G.821
G.826
,
,