7. Record the 1 kHz absolute errors at each output level from Table 22 into the
appropriate column in Table 24. Record the flatness errors from Table 23 for
each output level and frequency into the appropriate column in Table 24. Add
the errors and verify that they are within spec for the appropriate time
internal.
Table 21. Wideband Frequency Accuracy Test Record
Frequency ( Hz )
10 Hz
100 Hz
300 Hz
500 Hz
800 Hz
900 Hz
1 kHz
1.19 kHz
2.2 MHz
3.5 MHz
3.8 MHz
10 MHz
20 MHz
30 MHz
Table 22. Wideband Accuracy at 1 kHz Test Record
Measured
Output
Error
Level
ppm
2.1 V
1.0 V
0.3 V
0.1 V
30 mV
10 mV
3 mV
1 mV
Frequency Measured
5790A
Spec
± ppm
1-Year Spec
(90-day)
± ppm
2238
24
2900
24
2833
38
3400
53
3333
115
3900
240
4833
643
5400
1720
Tolerance Limits
99.99 ms to 100.01 ms
9.999 ms to 10.001 ms
299.97 Hz to 300.03 Hz
499.95 Hz to 500.05 Hz
799.92 Hz to 800.08 Hz
899.91 Hz to 900.09 Hz
999.0 Hz to 1.0001 kHz
1.189881 kHz to 1.190119 kHz
2.19978 MHz to 2.20022 MHz
3.49965 MHz to 3.50035 MHz
3.79962 MHz to 3.80038 MHz
9.990 MHz to 10.001 MHz
19.998 MHz to 20.002 MHz
29.997 MHz to 30.003 MHz
TUR
GB Factor
93
1.000
121
1.000
74.5
1.000
64.1
1.000
29.0
1.000
16.2
1.000
7.5
1.000
3.14
0.980
Multifunction Calibrator
Wideband Output Flatness Test
Test Limit
2238
2900
2833
3400
3333
3900
4833
5292
85