Fluke 5730A Calibration Manual page 70

Multifunction
Hide thumbs Also See for 5730A:
Table of Contents

Advertisement

5730A
Calibration Manual
60
5. Using a DMM does take longer and more care to self-
calibrate a 752A then using a Fluke 845 but it can still be
done in a reasonable amount of time once the user gets use
to the new technique. The user needs to determine which
settings of the 8508A work best for them. Settings such as
the number of digits, 7 or 8, using FAST or not or using the
FILTER or not, can be tried until a combination is found that
works best. It may be that one set of settings is found to be
best for the initial adjustments and then a change in the
settings for the final adjustment. For example, it may work
best to set up the meter to respond quickly to a change in
voltage to get the adjustments close then change to a setting
that takes longer per reading but has less scatter for the final
readings.