Table 9. DC Voltage Test at Decade Values Verification
10 V
Divider
Product
Std
Setting
Range
Polarity
POS
0.1
220 mV
NEG
0.1
220 mV
POS
1
2.2 V
NEG
1
2.2 V
POS
10
11 V
NEG
10
11 V
POS
10
22 V
NEG
10
22 V
POS
100
220 V
NEG
100
220 V
POS
100
1100 V
NEG
100
1100 V
POS
1000
1100 V
NEG
1000
1100 V
Vstd=
DMM
Product
Reading
Output
(Vdmm)
V
100 mV
100
-100 mV
V
1 V
-1 V
V
10 V
-10 V
V
10 V
-10 V
(
100 V
(
-100 V
(
100 V
(
-100 V
(
1000 V
(
-1000 V
Calculated
Equation
Measured
Voltage
std
V
dmm
V
std
V
dmm
100
std
V
dmm
10
V
std
V
dmm
10
+
V
std
dmm
+
V
V
std
dmm
+
V
std
dmm
+
V
V
std
dmm
)
+
V
V
10
std
dmm
)
+
V
V
10
std
dmm
)
+
V
V
10
std
dmm
)
+
V
V
10
std
dmm
)
+
V
V
100
std
dmm
)
+
V
V
100
std
dmm
Multifunction Calibrator
Full Verification
Limits
±1.2 μV
±1.2 μV
±4.8 μV
±4.8 μV
±33 μV
±33 μV
±35 μV
±35 μV
±450 μV
±450 μV
±1.1 mV
±1.1 mV
±6.5 mV
±6.5 mV